3 Operation Model 6000A
3-30 Part I
Teledyne Analytical Instruments
Before starting linearization, perform a standard calibration. See section
4.4. To enter data:
From the
System
Functions screen—
1. Use < > to select ALGORITHM , and
Enter
.
2. Select and
Enter
SETUP.
3.
Enter
AUTO from the Calibration Mode Select screen.
The Auto Linearize Mode data entry screen appears.
19.5 ppm SO2
Input(Ø) :20.00
5. Use the ∆∇ keys to set the proper value of calibration gas, and
Enter
. Repeat this step for each cal-point number as it appears in
the Input (x) parentheses.
6. Repeat step 5 for each of the special calibration gases, from the
lowest to the highest concentrations. Press
Escape
when done.
To end the session, send:
st<enter>
st<enter>
to the analyzer from the computer.