User guide
Setup Menu Model 6200T Total Sulfur Analyzer
Teledyne Analytical Instruments 104
milligrams per cubic meter (mg/m
3
, MGM). Changing units affects all of the display,
analog outputs, COM port and DAS values for all reporting ranges regardless of the
analyzer’s range mode.
To change the concentration units:
SETUP X.X CONC UNITS: PPM
PPM PPB UGM MGM ENTER EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP X.X CONC UNITS: PPB
PPM PPB UGM MGM ENTER EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
Select the preferred
concentration unit.
EXIT returns
to the main menu.
ENTR accepts
the new unit,
EXIT returns
to the SETUP
menu.
Figure 5-7: SETUP RNGE – Concentration Units Selection
IMPORTANT
IMPACT ON READINGS OR DATA
Concentrations displayed in mg/m
3
and µg/m
3
use 0°C and 760 Torr as
standard temperature and pressure (STP). Consult your local regulations
for the STP used by your agency. Here are the conversion factors from
volumetric to mass units used in the 6200T: SO
2
: ppb x 1.34 = µg/m
3
; ppm
x 1.34 = mg/m
3