6200T Total Sulfur Analyzer Setup Menu
Teledyne Analytical Instruments 137
5.9.9. Test Channel Output
When activated, output channel A3 can be used in the standard configuration to report
one of the test functions viewable from the SAMPLE mode display.
To activate the A3 channel and select a test function, access the Signal I/O from the
DIAG Menu (refer to Figure 5-16), then press:
DIAG TCHN TEST CHANNEL: NONE
NEXT ENTR EXIT
DIAG TEST CHAN OUTPUT
PREV NEXT ENTR EXIT
Continue to press NEXT until …
DIAG TCHN TEST CHANNEL: PMT READING
PREV NEXT ENTR EXIT
Press PREV or NEXT
to move through the
list of available
parameters
(Table 6-9)
Press EXIT to
return to the
DIAG menu
Press ENTR to
select the displayed
parameter and to
activate the test
channel.
DIAG SIGNAL I / O
PREV NEXT ENTR EXIT
Figure 5-35: DIAG – Test Channel Output
Table 5-9: Test Parameters Available for Analog Output A3 (standard
configuration)
Test Channel Test parameter range
NONE
Test channel is turned off
PMT READING
0-5000 mV
UV READING
0-5000 mV
SAMPLE PRESSURE
0-40 in-Hg-A
SAMPLE FLOW
0-1000 cm³/min
RCELL TEMP
0-70° C
CHASSIS TEMP
0-70° C
IZS TEMP
0-70° C
PMT TEMP
0-50° C
HVPS VOLTAGE
0-5000 V