User guide
6200T Total Sulfur Analyzer Communications Setup
Teledyne Analytical Instruments 147
6.5.2. Configuring Ethernet Communication Using Dynamic Host
Configuration Protocol (DHCP)
1. Consult with your network administrator to affirm that your network server is running
DHCP.
2. Access the Communications Menu as shown in Figure 5-13.
3. Follow the setup sequence as shown in Figure 6-5.
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
SETUP X.X INST IP: 0.0.0.0
<SET SET> EXIT
SETUP X.X GATEWAY IP: 0.0.0.0
<SET SET> EXIT
SETUP X.X TCP PORT2: 502
<SET SET> EDIT EXIT
EDIT button
disabled
From this point on,
EXIT returns to
COMMUNICATIONS
MENU
Do not alter unless
directed to by Teledyne
Instruments Customer
Service personnel
SETUP X.X HOSTNAME:
<SET EDIT EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X SUBNET MASK: 0.0.0.0
<SET SET> EXIT
SETUP X.X TCP PORT: 3000
<SET SET> EDIT EXIT
SETUP X.X DHCP: ON
SET> EDIT EXIT
DHCP: ON is
default setting.
If it has been
set to OFF,
press EDIT
and set to ON.
SETUP X.X DHCP: OFF
OFF ENTR EXIT
SETUP X.X DHCP: ON
ON ENTR EXIT
Figure 6-5: COMM – LAN / Internet Automatic Configuration
6.5.2.1. Changing the Analyzer’s HOSTNAME
The HOSTNAME is the name by which the analyzer appears on your network. The
default name for all Teledyne Analytical Instruments’s 6200T analyzers is 6200T. To