6200T Total Sulfur Analyzer Calibration Procedures
Teledyne Analytical Instruments 211
9.10.2.3. Activate CO
2
Sensor Stability Function
To change the stability test function from SO
2
concentration to the CO
2
sensor output,
press:
Press 3
times to return
to
menu.
SETUP X.X
SO2 CO2 O2 ENTR
SAMPLE RANGE=50.0 PPM SO2= XX.XX
< TST TST > CAL
SETUP X.X
CFG DAS RNGE PASS CLK EXIT
SETUP X.X
COMM DIAG ALRM EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV JUMP EDIT PRNT EXIT
SETUP X.X
<PREV NEXT> JUMP PRNT EXIT
SETUP X.X
SO2 O2 ENTR EXIT
Continue pressing until ...
SETUP X.X
EXIT
CO
2
and O
2
options only
appear if
associated
sensors are
installed.
Figure 9-19: Activate CO
2
Sensor Stability Function
IMPORTANT
IMPACT ON READINGS OR DATA
Use the same procedure to reset the STB test function to SO
2
when the
CO
2
calibration procedure is complete.