6200T Total Sulfur Analyzer Maintenance
Teledyen Analytical Instruments 235
9. Wait 5 minutes.
If the gauge pressure drops >1 psi within 5 minutes, then the hydrocarbon scrubber has
an internal leak and must be replaced. Contact Teledyne Analytical Instruments’s
Customer Service.