12.7.2. Sensor Module Repair & Cleaning
CAUTION - GENERAL SAFETY HAZARD
Do not look at the UV lamp while the unit is operating. UV light can cause
eye damage. Always use safety glasses made from UV blocking material
when working with the UV Lamp Assembly. (Generic plastic glasses are
not adequate).
Figure 12-6: Sensor Module Wiring and Pneumatic Fittings
IMPORTANT
IMPACT ON READINGS OR DATA
After any repair or service has been performed on the sensor module, the
6200T should be allowed to warm up for 60 minutes.
Always perform a leak check (refer to Section 11.3.6) and calibrate the
analyzer (refer to Section 9) before placing it back in service.