User guide

Instrument Test and Calibration Record Model 6200T Total Sulfur Analyzer
342 Teledyne Analytical Instruments
Table 17-2.
For test and calibration information with CO2, refer to Table 17-3.
Table 17-1. Final Test and Calibration Values for 6200T
TEST Parameters Observed Value Units Acceptable Value
RANGE PPB 50 - 20,000
STABIL PPB 0.0 - 2
PRESS “ HG 24 - 35
SAMP FL CC / MIN 500 - 700 w/CO
2
PMT mV 0 - 5000
UV LAMP mV 3500 - 4000
STR. LGT PPB < 60
DRK PMT MV < 50
DRK LMP MV < 50
SLOPE 1.0 ± 0.3
OFFSET MV < 100
HVPS V 400 - 900 constant
DCPS MV 2500 +/- 200
RCELL TEMP
o
C 50 +/- 1
BOX TEMP
o
C 8-50
PMT TEMP
o
C 7.9 +/- 1
IZS TEMP
o
C 50 +/-.3
Electric Test
PMT Volts MV 1000 +/-200
TS Conc PPB 500 +/- 100
Optic Test
PMT Volts MV 1000 +/- 200
TS Conc PPB 500 +/- 100