User guide

6200T Total Sulfur Analyzer Instrument Test and Calibration Record
Teledyen Analytical instruments 343
Table 17-2. Test and Calibration Values for 6200TU
TEST
PARAMETERS
OBSERVED
VALUE
UNITS ACCEPTABLE VALUE(S)
RANGE
PPB
5 - 20,000
STAB1
PPB
0.05 ppb with zero air
STAB2
PPB
0.1 ppb with zero air
PRESS HG
ambient ± 2
SAMPLE FL CC / MIN
650 cm3/min ± 10%
PMT mV -20 TO 150 mV with zero air
UV LAMP mV 2000 - 4800
STR LGT PPB < 25
DRK PMT MV 200 - 325
DRK LMP MV -50 - 200
SLOPE 1.0 ± 0.3
OFFSET MV < 250
HVPS V 400 to 900
RCELL TEMP
o
C 50 ± 1º
BOX TEMP
o
C ambient
+ ~ 5
PMT TEMP
o
C 7 ± 2
IZS TEMP (option)
o
C 50 ± 1
Electric Test
PMT Volts MV 1000 +/-200
TS Conc PPB 500 +/- 100
Optic Test
PMT Volts MV 1000 +/- 200
TS Conc PPB 500 +/- 100