Instrument Test and Calibration Record Model 6200T Total Sulfur Analyzer
344 Teledyne Analytical Instruments
Table 17-3. Test and Calibrations Values w/ CO2 where applicable
Span and Cal Value
s
Acceptable Value
Parameter Observed Value Units Nominal Range
TS Span Conc. PPB 20 - 20,000
TS Slope 1.0 +/- .3
TS Offset MV < 100
Noise at Zero (rms) PPB < 0.2
Noise at Span (rms) PPB < 0.5
PMT at Zero
(SO
2
/CO
2
)
MV
PMT at Span
(SO
2
/CO
2
)
MV
Measured Flows
Parameter Observed Value Units Nominal Range
Sample Flow w/ CO
2
cc/min 500 - 700
Sample Flow w/Air cc/min 400 - 600
Sample Press w/CO
2
“ HG 24 - 27
IZS Purge Flow cc/min 50 +/- 10
H
2
S Conversion
Efficiency
Expected =
______
PPB
Actual = ______
PPB
Efficiency = ________ %
(100 2%)
Factory Installed Options Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Internal Zero/Span - IZS
Permeation Tube (Output Specification)
4-20 MA Current Loop Output
External Pump
PROM Rev #: _______________________ 6200TTS S/N: ____________
M501TS S/N: ____________
Date: _______________________ Technician: _______________________