ADDENDUM TO MODEL 100E OPERATORS MANUAL (P/N 04515) FOR MODEL 100EU TRACE LEVEL SULFUR DIOXIDE ANALYZER 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: Phone: Fax: Email: Website: © 2007 T-API PRINT DATE: 08 April 2009 800-324-5190 858-657-9800 858-657-9816 Api-sales@teledyne.com http://www.teledyne-api.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04515 TABLE OF CONTENTS TABLE OF CONTENTS 1. PREFACE .......................................................................................................................................................................... 3 1.1 Reference Numbering convention............................................................................................................................. 3 2.
TELEDYNE INSTRUMENTS TABLE OF CONTENTS M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 LIST OF FIGURES Figure 3-1: Figure 3-2: Figure 6-1: Figure 10-2: Figure 10-3: Figure 11-1: Figure 11-2: Figure 11-3: M100EU Internal Layout ................................................................................................................7 M100EU Test Functions ................................................................................................................
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS PREFACE Addendum to M100E Manual - P/N 04515 1. PREFACE NOTE The information contained in this addendum is pertinent to M100EU analyzers running software revision F.0. Some or all of the information may not be applicable to later revisions of software. The software revision your analyzer is running is displayed in the upper left-hand corner of the display any time the instrument is in SETUP mode.
TELEDYNE INSTRUMENTS PREFACE M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 USER NOTES: 4 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS SPECIFICATIONS, APPROVALS AND WARRANTY Addendum to M100E Manual - P/N 04515 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1 SPECIFICATIONS Table 2-1: Model 100EU Basic Unit Specifications Min/Max Range (Physical Analog Output) In 1ppb increments from 50ppb to 20 000ppb, dual ranges or auto ranging Measurement Units ppb, ppm, µg/m3, mg/m3 (user selectable) 2 Zero Noise 25 ppt RMS (50 ppt RMS with 360 nM filter installed) 2 Span Noise 0.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL SPECIFICATIONS, APPROVALS AND WARRANTY Addendum to M100E Manual - P/N 04515 2.2 EPA EQUIVALENCY DESIGNATION At this time the M100EU has not been certified by the EPA as an equivalent method at the time of this writing however it is anticipated that the M100EU will qualify as Reference Method Number EQSA-0495-100 per 40 CFR Part 53 in the near future. Please see section 2.2 of the M100E manual, P/N 04515 for details. 2.3 CE MARK COMPLIANCE See Section 2.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS GETTING STARTED Addendum to M100E Manual - P/N 04515 3. GETTING STARTED 3.1 UNPACKING THE M100EU Unpack the M100EU as per the directions in Section 3.1 of the M100E Manual - P/N 04515, with the following change. There are two redheaded shipping screws that hold down the PMT/Sensor assembly and must be removed prior to operation. They are located along the base of the PMT housing adjacent to the chassis. 3.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL GETTING STARTED Addendum to M100E Manual - P/N 04515 3.3 FUNCTIONAL CHECK OF THE M100EU To perform an initial functional check of the M100EU follow the steps contained in Section 3.2.4 of the M100E Manual - P/N 04145, but use the Test functions described in below. 3.3.1 TEST FUNCTIONS Table 3-1 supersedes the figure in Step 2 of Section 3.2.4 and Figure 6-2 of the M100E Manual - P/N 04145. Table 11-1 supersedes Table 6-2.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS GETTING STARTED Addendum to M100E Manual - P/N 04515 Table 3-1: Test Functions Defined DISPLAY PARAMETER UNITS OFFSET SO2 measurement Offset - The overall offset of the instrument as calculated during the last calibration activity. The offset parameter is used to set the zero point of the analyzer response. HVPS HVPS V The PMT high voltage power supply. RCELL TEMP SAMPLE CHAMBER TEMP °C The current temperature of the sample chamber.
TELEDYNE INSTRUMENTS GETTING STARTED M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 USER NOTES: 10 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 TELEDYNE INSTRUMENTS FREQENTLY ASKED QUESTIONS & GLOSSARY 4. FREQENTLY ASKED QUESTIONS & GLOSSARY Please refer to Section 4 of the M100E manual, PN 04515 for information.
TELEDYNE INSTRUMENTS FREQUENTLY ASKED QUESTIONS & GLOSSARY M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 USER NOTES: 12 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 TELEDYNE INSTRUMENTS OPTIONAL HARDWARE AND SOFTWARE 5. OPTIONAL HARDWARE AND SOFTWARE Please refer to Section 5 of the M100E manual, PN 04515 for information.
TELEDYNE INSTRUMENTS OPTIONAL HARDWARE AND SOFTWARE M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 USER NOTES: 14 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS M100EU OPERATING INSTRUCTIONS Addendum to M100E Manual - P/N 04515 6. M100EU OPERATING INSTRUCTIONS NOTE For the most part the operation instruction for the M100EU are the same as those described in Chapter 6 of the M100E Manual - P/N 04515 with the exception that there are additional test parameters and setup procedures. 6.1 ADDITIONAL TEST PARAMETERS Please see Section 3.3 above for details on the additional test parameters. 6.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL M100EU OPERATING INSTRUCTIONS Addendum to M100E Manual - P/N 04515 SAMPLE RANGE = 500.000 PPB SO2 =XXX.X SETUP < TST TST > CAL SETUP X.X EXIT returns to the main SAMPLE display CFG DAS RNGE PASS CLK MORE SETUP X.X EXIT returns to the PRIMARY SETUP MENU EXIT SECONDARY SETUP MENU COMM VARS DIAG STBL SETUP X.X NEXT SETUP X.X CONC2 is setup for STABIL2 test function, UVLAMP for UV STAB test function PRIMARY SETUP MENU PREV NEXT SETUP X.
M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04145 TELEDYNE INSTRUMENTS CALIBRATION PROCEDURES 7. CALIBRATION PROCEDURES Calibration of the M100EU should be performed according to the procedures described in Chapters 7 & 8 of the M100E Manual - P/N 04145. However, delivering span and zero gases for the lower ranges that the M100EU is designed for can be difficult.
TELEDYNE INSTRUMENTS CALIBRATION PROCEDURES M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 USER NOTES: 18 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 EPA PROTOCOL CALIBRATION 8. EPA PROTOCOL CALIBRATION The M100EU is designated as an equivalent method for measuring SO2 under certification EQSA-0495-100. Calibration of the M100EU is done so in a manner that is consistent with EPA requirements. For calibration in this manner please refer to section 8 of the M100E manual.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL EPA PROTOCOL CALIBRATION Addendum to M100E Manual - P/N 04515 USER NOTES: 20 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04145 TELEDYNE INSTRUMENTS INSTRUMENT MAINTENANCE 9. INSTRUMENT MAINTENANCE Instrument maintenance is almost identical to that in the M100E. The M100EU uses a 1 micron sample filter, instead of the 5 micron sample filter used in the M100E. Replacement part numbers are shown below.
TELEDYNE INSTRUMENTS INSTRUMENT MAINTENANCE M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 USER NOTES: 22 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS THEORY OF OPERATION Addendum to M100E Manual - P/N 04145 10. THEORY OF OPERATION The M100EU is a modified M100E. The primary differences are the way in which the PMT and UV reference signals are acquired and processed. The M100EU has no shutter but rather employs synchronous demodulation to capture the dark and light PMT and UV reference signals several times per second.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL THEORY OF OPERATION Reaction Cell Addendum to M100E Manual - P/N 04515 Lamp Shutter Housing UV LAMP Sample Gas Outlet PMT PREAMP PCA SAMPLE CHAMBER Sample Gas Inlet PMT HOUSING Reference Detector PMT Cooling System Figure 10-2: M100EU Sensor Module Assembly These components are divided into two significant subassemblies. The sample chamber and the PMT assembly.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS THEORY OF OPERATION Addendum to M100E Manual - P/N 04145 UV Source Lamp Shutter Housing UV Source Lens & Housing Sample Air Outlet O-Ring Seal PMT Lens & Housing O-Ring Seal Sample Chamber Heater 1 Sample Air Inlet Sample Chamber Sample Chamber Temperature Sensor & Heater 2 Figure 10-3: 05944 Rev B DCN 5063 Heater Safety Shut-Off Light Trap Reference Detector M100EU Sample Chamber Exploded View PRINTED DOCUMENTS ARE UNCONTROLLED 25
TELEDYNE INSTRUMENTS THEORY OF OPERATION M100EU – OPERATIONS MANUAL Addendum to M100E Manual - P/N 04515 User Notes: 26 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 TROUBLESHOOTING & REPAIR 11. TROUBLESHOOTING & REPAIR This section includes various troubleshooting and repair information that is either in addition to that included in Chapter 11 of the M100E Manual (P/N 04145) CAUTION THE OPERATIONS OUTLINED IN THIS CHAPTER MUST BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL TROUBLESHOOTING & REPAIR Addendum to M100E Manual - P/N 04515 Table 11-1: Test Functions - Possible Causes for Out-Of-Range Values TEST FUNCTION NOMINAL VALUE(S) NORM PMT -- POSSIBLE CAUSE(S) Noisy Norm PMT value (assuming unchanging SO2 concentration of sample gas): Calibration error; HVPS problem; PMT problem; UV reference problem; UV lamp problem. This is the instantaneous reading of the UV lamp intensity.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 TROUBLESHOOTING & REPAIR 11.1.3 FAULT DIAGNOSIS WITH SYNC DEMOD PCA LEDS There are four green Light Emitting Diodes (LEDs) on the bottom left side of the Sync Demodulator PCA. They indicate various statuses and can be used to troubleshoot problems associated with the board.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL TROUBLESHOOTING & REPAIR Addendum to M100E Manual - P/N 04515 The UV output level of the lamp is not even across the entire length of the lamp. Some portions of the lamp shine slightly more brightly than others. At the factory the position of the UV lamp is adjusted to optimize the amount of UV light shining through the UV filter/lens and into the reaction cell.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 TROUBLESHOOTING & REPAIR Figure 11-1: UV Lamp Adjustment 4.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL TROUBLESHOOTING & REPAIR Addendum to M100E Manual - P/N 04515 11.3.1.2 Replacing the UV Lamp 1. Turn off the analyzer. 2. Disconnect the UV lamp from its power supply. You can find the power supply connector by following the two, white UV Lamp power supply wires from the lamp to the power supply. 3.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 TROUBLESHOOTING & REPAIR 11.3.2 REPLACING THE UV FILTER/LENS Note: Be careful not to leave thumbprints on the interior of the sample chamber. The various oils that make up fingerprints fluoresce brightly under UV light and will significantly affect the accuracy of the analyzer’s SO2 measurement) 1. Turn off the instrument’s power and remove the power cord from the instrument. 2.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL TROUBLESHOOTING & REPAIR Addendum to M100E Manual - P/N 04515 11.3.3 REPLACING THE PMT, HVPS OR TEC The PMT should last for the lifetime of the analyzer. However, in some cases, the high voltage power supply (HVPS) or the thermo-electric cooler (TEC) may fail.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 TROUBLESHOOTING & REPAIR Because the threads of the plastic screws are easily damaged it is highly recommended to use new screws when reassembling the unit. 9. Carefully take out the assembly consisting of the HVPS, the gasket, preamp assembly and the PMT. 10. Change the PMT or the HVPS or both, clean the PMT glass tube with a clean, anti-static wipe and do not touch it after cleaning. 11.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL TROUBLESHOOTING & REPAIR Addendum to M100E Manual - P/N 04515 11.3.4 M100EU PMT HARDWARE CALIBRATION (FACTORY CAL) THIS PROCEDURE SUPERSEDES THE ONE CONTAINED IN SECTION 11.6.3 OF THE M100E MANUAL P/N 04145. The sensor module hardware calibration adjusts the slope of the PMT output when the Instruments slope and offset values are outside of the acceptable range and all other more obvious causes for this problem have been eliminated.
M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS Addendum to M100E Manual - P/N 04145 TROUBLESHOOTING & REPAIR 11.
TELEDYNE INSTRUMENTS M100EU – OPERATIONS MANUAL TROUBLESHOOTING & REPAIR Addendum to M100E Manual - P/N 04515 USER NOTES: 38 PRINTED DOCUMENTS ARE UNCONTROLLED 05944 Rev B DCN 5063
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software Documentation APPENDIX A-1: MODEL 100EU SOFTWARE MENU TREES APPENDIX A-2: MODEL 100EU SETUP VARIABLES AVAILABLE VIA SERIAL I/O APPENDIX A-3: MODEL 100EU WARNINGS AND TEST MEASUREMENTS VIA SERIAL I/O APPENDIX A-4: MODEL 100EU SIGNAL I/O DEFINITIONS APPENDIX A-5: MODEL 100EU IDAS FUNCTIONS APPENDIX A-6: MODEL 100EU TERMINAL COMMAND DESIGNATORS 05928 Rev F.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100EU Software Menu Trees, Revision F.0B Model 100EU Instruction Manual APPENDIX A-1: M100E Software Menu Trees, Revision F.0B Pg: A-2 PRINTED DOCUMENTS ARE UNCONTROLLED 05928 Rev F.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100E Software Menu Trees, Revision F.0B Model 100E Instruction Manual SAMPLE TEST 1 Only appear if reporting range is set for A UTO range m ode. RANGE STABIL STABIL2 PRES SAMP FL PMT NOR M PMT UV LAMP UV STB LAMP RATIO STR.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100EU Software Menu Trees, Revision F.0B Model 100EU Instruction Manual SAMPLE TEST1 Only appear if reporting range is set for AUTO range mode. LOW RANGE STABIL ZERO STABIL2 PRES SAMP FL PMT NORM PMT UV LAMP LAMP RATIO STR.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100E Software Menu Trees, Revision F.0B Model 100E Instruction Manual SETUP ENTER SETUP PASS: 818 CFG PREV DAS ACAL1 NEXT NEXT MODE SET2 TIME CONFIGURATION SAVED ENTR 2 3 Only appears if a applicable option/feature is installed and activated. Only appears whenever the currently displayed sequence is not set for DISABLED. Only appears when reporting range is set to AUTO range mode.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100EU Software Menu Trees, Revision F.0B Model 100EU Instruction Manual SAMPLE ENTER SETUP PASS: 818 CFG DAS ACAL1 RNGE VIEW PREV CLK MORE EDIT NEXT CONC PNUMTC CALDAT DETAIL FAST PREV PRM> Cycles through lists of parameters chosen for this iDAS channel PV10 PREV NEXT NX10 PREV INS DEL YES EDIT SET> EDIT PRNT Creates/changes name NAME EVENT PARAMETERS REPORT PERIOD NUMBER OF RECORDS RS-232 REPORT CHANNEL ENABLE CAL.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100E Software Menu Trees, Revision F.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100EU Software Menu Trees, Revision F.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100E Software Menu Trees, Revision F.0B Model 100E Instruction Manual SETUP . ENTER SETUP PASS: 818 CFG DAS ACAL1 RNGE PASS COMM HESN2 ID PREV RESPONSE MODE BCC NEXT TEXT INS EDIT YES GAS LIST STATUS FLAGS CMD DEL EDIT PRNT Select from list of available gases (see Section 6.12.4.6). NO GAS TYPE GAS ID REPORTED SO2, 110, REPORTED ON OFF Figure A-7: 05928 Rev F.
TELEDYNE INSTRUMENTS APPENDIX A-1: M100EU Software Menu Trees, Revision F.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0B Table A-1: M100EU Setup Variables, Revision F.0B M100EU Setup Variables for Latest Revision Setup Variable Numeric Units DAS_HOLD_OFF Minutes TPC_ENABLE — ON, Default Value 15 OFF RCELL_SET ºC Value Range 0.5–20 Duration of DAS hold off period. OFF, ON ON enables temperature and pressure compensation; OFF disables it.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation sample. 0.2 4 FILT_SIZE Samples 240, 1–480 Moving average filter size. 1–100 Moving average filter size in adaptive mode. 0.001–0.1, Absolute change to trigger adaptive filter. 3 30 , 54 FILT_ASIZE Samples 20, 6 FILT_DELTA FILT_PCT 3 PPM 0.02, 10 3 10 2 3 1–100 % 5, FILT_DELAY Seconds 180 FILT_ADAPT — ON, OFF 1–100 Percent change to trigger adaptive filter.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation O2_FILT_DELTA 10 %2 0.1–100 Absolute change in O2 concentration to shorten filter. O2_FILT_PCT 10 %2 0.1–100 Relative change in O2 concentration to shorten filter. O2_FILT_DELAY 10 Seconds 20 0–300 Delay before leaving O2 adaptive filter mode. O2_DIL_FACTOR 10 —1 0.1–1000 Dilution factor for O2. Used only if is dilution enabled with FACTORY_OPT variable.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual DARK_MODE 5 APPENDIX A - Version Specific Software Documentation — AUT O OFF, AUTO, CMND DARK_ENABLE 1, 3, 4 — ON, OFF DARK_FREQ OFF, ON ON enables PMT/UV dark calibration; OFF disables it. 0.1–1440 Dark calibration period. 1–60 Dwell time after closing dark shutter or turning off lamp or selecting preamp range. 1–180 Dwell time after opening dark shutter or turning on lamp. 4 Minutes 30, Dark calibration mode.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation AUTO2 PHYS_RANGE1 0.1–2500, PPM 2, 500 PHYS_RANGE2 3 PPM 22, CONC_RANGE1 0.1–2500, 5500 3 5000 3 Conc 500, CONC_RANGE2 Low pre-amp range. 5–10000 3 5–10000 High pre-amp range. 3 0.1–50000 D/A concentration range 1. 0.1–50000 D/A concentration range 2. % 15 0.1–500 CO2 concentration range. % 100 0.1–500 O2 concentration range.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual RS232_MODE APPENDIX A - Version Specific Software Documentation BitFlag 0, 0–65535 3 5 RS-232 COM1 mode flags. Add values to combine flags.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation 19200, 38400, 57600, 115200 MODEM_INIT2 — “AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0” Any character in the allowed character set. Up to 100 characters long. RS-232 COM2 modem initialization string. Sent verbatim plus carriage return to modem on power up or manually. Enclose value in double quotes (“) when setting from the RS-232 interface.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation RCELL_PROP 1/ºC RCELL_INTEG 0–10 Reaction cell temperature PID proportional coefficient. — 0.005 0–10 Reaction cell temperature PID integral coefficient. RCELL_DERIV — 0.5 0–10 Reaction cell temperature PID derivative coefficient. O2_CELL_CYCLE 10 Seconds 10 0.5–30 O2 cell temperature control cycle period.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation DIM I2C_RESET_ENABLE — ON CLOCK_FORMAT — “TIME=%H:% M:%S” OFF, ON 2 I C bus automatic reset enable. Any character in the allowed character set. Up to 100 characters long. Time-of-day clock format flags. Enclose value in double quotes (“) when setting from the RS-232 interface. “%a” = Abbreviated weekday name. “%b” = Abbreviated month name. “%d” = Day of month as decimal number (01 – 31).
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation 2048 = enable Internet option 4096 = enable pre-amplified UV lamp monitoring 1 M100E. 2 M100ES. 3 M100E 4 M100EF 5 RP84 6 7 H. . 00S. Engineering firmware only. iChi 8 p option. Must power-cycle instrument for these options to fully take effect. 9 Low span option. 10 O2 option. 11 CO 12 M100E U. 99 Obsolete. A-20 2 option. PRINTED DOCUMENTS ARE UNCONTROLLED 05928 Rev F.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A-3: Warnings and Test Functions, Revision F.0B Table A-2: M100EU Warning Messages, Revision F.0B Name 1 WSYSRES SYST WDATAINIT WCONFIGINIT CONFIG Message Text EM RESET DATA INITIALIZED INITIALIZED Description Instrument was power-cycled or the CPU was reset. Data storage was erased. Configuration storage was reset to factory configuration or erased.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation Table A-3: Name 1 = RANGE RANGE M100EU Test Functions, Revision F.0B Message Text 500.0 PPB 3 Description D/A range in single or auto-range modes. SO2 RNG=500.0 PPB 3, 10, 11 = RANGE1 RANGE1 50 0.0 PPB 3 D/A #1 range in independent range mode. SO2 RN1=500.0 PPB 3, 10, 11 = RANGE2 RANGE2 50 0.0 PPB 3 SO2 RN2=500.0 PPB CO2 RNG=100 PCT 11 CO CO2RANGE D/A #2 range in independent range mode.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation calibration. HVPS HVPS=650 VOLTS High voltage power supply output. RCELLDUTY RCELL ON=0.00 SEC Reaction cell temperature control duty cycle. RCELL TEMP=52.1 C Reaction cell temperature. O2 CELL TEMP=50.2 C O2 sensor cell temperature. BOXTEMP BOX TEMP=35.5 C Internal chassis temperature. PMTTEMP PMT TEMP=7.0 C PMT temperature. IZSDUTY IZS ON=0.00 SEC IZS temperature control duty cycle.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A-4: M100Eu Signal I/O Definitions, Revision F.0B Table A-4: M100EU Signal I/O Definitions, Revision F.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation Control outputs, U21, J1008, pins 9–12 = bits 0–3, default I/O address 325 hex 0–3 Spare Alarm outputs, U21, J1009, pins 1–12 = bits 4–7, default I/O address 325 hex ST_SYSTEM_OK2, MB_RELAY_36 4 9 1 = system OK 0 = any alarm condition or in diagnostics mode Controlled by MODBUS coil register MB_RELAY_37 9 5 Controlled by MODBUS coil register MB_RELAY_38 9 6 Controlled by MODBUS coil registe
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation MAINT_MODE 5 (input) 0 = maintenance mode LANG2_SELECT 6 (input) 0 = select second language 1 = normal mode 1 = select first language (English) SAMPLE_LED 8 (output) 0 = sample LED on 1 = off CAL_LED 9 (output) 0 = cal.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation AGND_DARK 3 7 AGND reading during dark cycles 3 8 AGND reading during light cycles 9 VREF4096 reading during dark cycles 10 VREF4096 reading during light cycles AGND_LIGHT VREF_DARK 3 VREF_LIGHT 3 Rear board primary MUX analog inputs PMT detector HVPS_VOLTAGE 1 HV power supply output PMT_TEMP 2 PMT temperature UVLAMP_SIGNAL 3 UV lamp intensity 4 PHOTO_ABS Temperature MUX 8 5 Pre
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual TEST_OUTPUT, APPENDIX A - Version Specific Software Documentation 3 DATA_OUT_4 Test measurement output, Data output #4 I2C analog output (AD5321), default I2C address 18 hex LAMP_POWER 5 1 Optional. 2 M100E H. 3 M100E U. 4 M100EF Engineering firmware only. 6 Low span option. 7 Pressurized span option. M100E. 9 MODBUS 10 11 Lamp power (0–5V) . 5 8 0 option. O2 option. CO A-28 2 option.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation M100E MODBUS Register Map MODBUS Register Address (dec.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation 102 10 O2 concentration during zero/span calibration, just before computing new slope and offset % 104 10 O2 slope — 106 10 O2 offset % 108 10 O2 sensor cell temperature C 200 11 202 11 CO 206 concentration CO2 concentration during zero/span calibration, just before computing new slope and offset 204 11 CO 11 2 CO % % 2 slope — 2 offset % MODBUS Floating Point Holding R
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual 100 10 In 101 10 200 11 APPENDIX A - Version Specific Software Documentation O2 calibration mode O2 cell temperature warning In CO2 calibration mode MODBUS Coil Registers (single-bit; read/write) 0 Maps to relay output signal 36 (MB_RELAY_36 in signal I/O list) 1 Maps to relay output signal 37 (MB_RELAY_37 in signal I/O list) 2 Maps to relay output signal 38 (MB_RELAY_38 in signal I/O list) 3 Maps to relay output signal 39 (MB_RELAY_39 in s
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A-5: M100EU iDAS Functions, Revision F.0B Table A-5: M100EU DAS Trigger Events, Revision F.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation SLOPE2 SO2 slope for range #2 — OFSET1 SO2 offset for range #1 mV OFSET2 SO2 offset for range #2 mV CO2SLP 11 CO2 slope — CO2OFS 11 CO2 offset % O2SLPE 10 O2 slope — O2OFST 10 O2 offset % SO2 concentration for range #1 during zero/span calibration, just before computing new slope and offset PPB, ZSCNC2 SO2 concentration for range #2 during zero/span calibration, just before c
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual 1 M100E. 2 M100E H. 3 M100E U. 10 APPENDIX A - Version Specific Software Documentation O2 option. 11 CO 2 option. 12 SO 2 with O2 correction option. A-34 PRINTED DOCUMENTS ARE UNCONTROLLED 05928 Rev F.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A-6: Terminal Command Designators, Revision F.0B Table A-7: Terminal Command Designators, Revision F.
TELEDYNE INSTRUMENTS Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation The command syntax follows the command type, separated by a space character. Strings in [brackets] are optional designators. The following key assignments also apply.
MODEL M100EU INSTRUCTION MANUAL Appendix B - Spare Parts Lists APPENDIX B - SPARE PARTS LISTS • 053600100 – M100EU Spare Parts List • 059460000 – M100EU Recommended Spares Stocking Levels 05929 Rev A B-1
Appendix B - Spare Parts Lists B-2 MODEL M100EU INSTRUCTION MANUAL 05929 Rev.
M100EU Spare Parts List Part Number 000940100 000940800 002690000 002700000 002720000 003290000 005960000 009690200 009690300 013140000 013210000 013390000 013400000 013420000 013570000 014400100 014750000 016290000 016300700 037860000 040010000 040030100 041710000 042410200 042580000 042900100 043570000 043940000 044670000 045150100 045230200 046250000 046260000 048620200 048830000 049310100 050610100 050610200 050610300 050610400 050630200 051990000 052660000 054340000 054650000 055100200 Description ORI
M100EU Spare Parts List Part Number 055560000 055920000 055930100 056080000 058020100 059200000 061930200 CN0000458 CN0000520 DS0000025 FL0000001 FL0000003 FM0000004 HW0000005 HW0000020 HW0000036 KIT000093 KIT000095 KIT000219 KIT000253 KIT000254 OP0000031 OR0000001 OR0000004 OR0000006 OR0000007 OR0000015 OR0000016 OR0000025 OR0000027 OR0000039 OR0000046 OR0000083 OR0000084 OR0000094 PU0000022 RL0000015 SW0000051 SW0000059 WR0000008 Description ASSY, VALVE, VA59 W/DIODE, 5" LEADS ASSY, SO2 SENSOR, M100EU (K
Teledyne Advanced Pollution Instrumentation RECOMMENDED SPARE PARTS STOCKING LEVELS Model 100EU PART NO 000940800 002720000 013140000 013400000 014080100 014610000 040010000 040030100 041710000 042410400 042580000 045230200 050630200 054340000 054650000 054710000 055120200 057020100 DS0000025 Orifice, 12 Mil Filter, 330 NM Cooler Fan PMT, SO2 Assy, HVPS, NOX/SOX Kit, Replacement Cooler Assy, Fan, Rear Panel, E Series PCA, Press Sensors (1X), Flow, E Series CPU, Configuration E Series Assy, Pump, Internal,
THIS PAGE IS INTENTIONALLY LEFT BLANK
Model 100EU Instruction Manual Addendum Appendix C – Warranty Repair Questionnaire Appendix C – Warranty Repair Questionnaire CUSTOMER: _________________________________ PHONE: __________________________________________________ CONTACT NAME: ____________________________ FAX NO. __________________________________________________ SITE ADDRESS: _________________________________________________________________________________________ MODEL 100EU SERIAL NO.
Model 100EU Instruction Manual Addendum 3. WHAT IS THE SAMPLE FLOW & SAMPLE PRESSURE W/SAMPLE INLET ON REAR OF MACHINE CAPPED? SAMPLE FLOW - 4. Appendix C – Warranty Repair Questionnaire CC SAMPLE PRESS - IN-HG-A WHAT ARE THE FAILURE SYMPTOMS? ______________________________________________________________ _______________________________________________________________________________________________________ 5. IF POSSIBLE, PLEASE INCLUDE A PORTION OF A STRIP CHART PERTAINING TO THE PROBLEM.
M100EU Manual Addendum Appendix D – Schematic List Appendix D – Schematic List Table D-1: List of Included Electronic Schematics DOCUMENT # 0506402 DOCUMENT TITLE M100EU UV Ref PCA 05435 M100EU Internal PMT Preamp PCA 05466 M100EU Sync/Demod PCA 0591602 M100EU Bursting UV Lamp Driver Other schematics can be found in the M100E Operators Manual (04145).
Appendix D – Schematic List D-2 M100EU Manual Addendum 05931 Rev.