APPENDIX A - Version Specific Software Documentation
Rev 1.0.3 (T-Series)/G6 (E-Series)
Teledyne API - Models T100, 100E Series (05036F DCN6650)
A-12 05036 Rev D
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
DWELL_TIME Seconds 1 0.1–10 Dwell time before taking each sample.
FILT_SIZE Samples 240,
30
3
,
480
12
1–1000 Moving average filter size.
FILT_ASIZE Samples 20,
6
3
1–100
Moving average filter size in adaptive
mode.
FILT_DELTA PPM 0.02,
10
3
,
0.005
12
0.001–0.1,
1–100
3
Absolute change to trigger adaptive
filter.
FILT_PCT % 5,
10
2
,
1
12
1–100
Percent change to trigger adaptive
filter.
FILT_DELAY Seconds 180 0–300
Delay before leaving adaptive filter
mode.
FILT_ADAPT — ON OFF, ON
ON enables adaptive filter; OFF
disables it.
NEG_CONC_SUPPRESS — OFF OFF, ON
ON pegs negative concentrations at
zero; OFF permits negative
concentrations
DIL_FACTOR — 1 0.1–1000
Dilution factor if dilution enabled with
FACTORY_OPT variable.
AGING_ENABLE
18
OFF ON, OFF ON enables aging offset and slope
compensation.
AGING_OFFSET_RATE
18
mV/day 0 -1.0–1.0 Aging offset rate of change per
day.
AGING_SLOPE_RATE
18
Change/
day
0 -0.01–0.01 Aging slope rate of change per
day.
CO2_DWELL
11
Seconds 1 0.1–30 Dwell time before taking each sample.
CO2_FILT_ADAPT
11
ON ON, OFF
ON enables CO
2
adaptive filter; OFF
disables it.
CO2_FILT_SIZE
11
Samples 48 1–300
CO
2
moving average filter size in
normal mode.
CO2_FILT_ASIZE
11
Samples 12 1–300
CO
2
moving average filter size in
adaptive mode.
CO2_FILT_DELTA
11
% 2 0.1–10
Absolute change in CO
2
concentration
to shorten filter.
CO2_FILT_PCT
11
% 10 0.1–100
Relative change in CO
2
concentration
to shorten filter.
CO2_FILT_DELAY
11
Seconds 90 0–300
Delay before leaving CO
2
adaptive filter
mode.
CO2_DIL_FACTOR
11
1 0.1–1000
Dilution factor for CO
2
. Used only if is
dilution enabled with FACTORY_OPT
variable.
CO2_STD_CELL_TEMP
11
ºK 323 1–500
Standard CO
2
cell temperature for
temperature compensation.
CO2_STD_CELL_PRESS
11
"Hg 28.50 1.00–50.00
Standard CO
2
cell pressure for
pressure compensation.
O2_DWELL
10
Seconds 1 0.1–30 Dwell time before taking each sample.
06807C DCN6650