Total Organic Carbon Analyzer Maintenance
Teledyne Analytical Instruments 55
5.3.5 CE Computer (P/N 13039)
If malfunctioning, replace the Computer Module using the
directions below. See Figure 5-7.
Figure 5-7: Replacing CE Computer
To Replace the CE Computer:
1. Turn off power
2. Disconnect electrical connectors.
3. Remove screws holding Module in place.
4. Remove module.
5. Install new CE Computer in reverse disassembly
instruction procedure steps. (Install electrical connectors
per supplied drawing).
6. Verify proper software install per directions shipped with
new CE Computer.
7. Recalibrate liquid zero and span values.