MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) MODEL GFC7000EU C O2 ANALYZER (Addendum to GFC7000E Manual PN 04584) Teledyne Analytical Instruments 16830 Chestnut St. City of Industry, Ca. 91748 TEL: FAX: TEL: E-MAIL: 626-961-9221 626-961-2538 626-934-1500 ai-customerservice@teledyne.com WEB SITE: http://www.teledyne-ai.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) OVERVIEW The model GFC7000EU is a close derivative of the Model GFC7000E CO Analyzer. The optical bench is longer (14 m), making the instrument more sensitive at low CO2 levels. The instrument has a user selectable full scale range of 0-100 PPB to 0-100 PPM of CO2. The instrument does not come equipped with a sample flow rate sensor. Hence, disregard any reference to the flow sensor in the users manual.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Humidity Range Temp Coefficient Voltage Coefficient Dimensions (HxWxD) Weight AC Power Environmental Conditions Analog Outputs Analog Output Ranges Analog Output Resolution Status Outputs Control Inputs I/O Certifications (1) (2) (3) (4) (5) 0-95% RH, Non-Condensing < 0.05 % of reading per °C (5 ppb/°C minimum)(6) < 0.05 % of reading per V 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm) 50 lb (22.7 kg) 100V 50/60 Hz (3.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2 exceeded. The relay alarm output AL1 is enabled all the time, whereas the AL2 and AL3 can be enabled/disabled by going into the diagnostics menu (with 929 password), then factory options, then turning conc. Alarm Relays ON or OFF.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) TEST deactivates Warning Messages until New warning(s) are activated SAMPLE WHEEL TEMP WARNING TEST CAL MSG RANGE=50.0 PPM SAMPLE CLR SETUP CO = XX.XX MSG < TST TST > CAL CO = XX.XX CLR SETUP MSG activates Warning Messages. keys replaced with TEST key All Warning messages are hidden, but MSG button appears WHEEL TEMP WARNING SAMPLE < TST TST > CAL MSG CO = XX.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Warning Message Fault Condition Possible Causes FRONT PANEL WARN The CPU is unable to Communicate with the Front Panel Display /Keyboard PHOTO TEMP WARNING PHT DRIVE is >4750 mVDC or <250 MVDC WARNING only appears on Serial I/O COM Port(s) Front Panel Display will be frozen, blank or will not respond.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Warning Message Fault Condition Possible Causes on wheel rotation. Failed Sync/Demod Board If Status LED’s on the Sync/Demod Board ARE flashing the cause is most likely an a problem with the analyzer’s main power: Intermittent loss of the power supply too short to cause a SYSTEM RESET warning; Frequency problem with the AC Mains. SYSTEM RESET The computer has rebooted.
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) #N05ATI 009690000 Filter, TFE, 47 mm 5 um, Qty 100 009690100 Filter, TFE, 47 mm 5 um, Qty 25 009820200 Assy, Motor, Synchronous, 115V/60HZ, "E" 010790000 Input Mirror 010800000 Output Mirror 016290000 Window, Sample Filter, 47 mm 016910000 Akit, Catalyst, CO/CO2 Converter (1 OZ) 019340000 Sample Thermistor M400 (S/N >099) or M401 (S/N>0573) 024710000 Tubing: 6', 1/8"CLR 024720000 Tubing: 6', 1/8" BLK
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) OR0000001 O-Ring, Flow Control OR0000058 O-Ring, Sample Filter, 47 mm, 2-228V OR0000034 O-Ring, Input/Output Mirror/Detector OR0000039 O-Ring, Window OR0000041 O-Ring, Field and Objective Mirror 043590100 Pump, 115V 50/60 Hz PU0000022 Pump Rebuild Kit, KNF Model #N05ATI PS0000025 Power Supply, +12V PS0000011 Power Supply, +5/+- 15V 042680000 Solenoid, Stainless Steel, 3-Way, 12V 042690000 Solenoid, S