The instrument’s zero air and span gas flow rate required for this option is 800 cc/min, however, the US EPA
recommends that the cal gas flow rate be at least 1600 cc/min.
Table 5-1: Zero/Span Valve Operating States
Option Mode Valve Condition
Sample/Cal Open to SAMPLE inlet
SAMPLE
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to ZERO/SPAN Valve
ZERO CAL
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to ZERO/SPAN Valve
50
SPAN CAL
Zero/Span Open to SPAN GAS inlet
The state of the Sample/Cal valves can be controlled:
Manually via the analyzer’s front panel;
By activating the instrument’s AutoCal
feature (See Section 9.4);
Remotely by using the External Digital I/O Control Inputs (See Section 9.3.2.3), or;
Remotely via the RS-232/485 Serial I/O ports (See Section 8.1.7).