Manual Addendum Model T360U CO2 Analyzer (Addendum to T360 Manual, PN 07272) © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.com Website: http://www.teledyne-api.
ABOUT THIS ADDENDUM This addendum, part number 07273, is to be used in conjunction with the Model T360 operation manual, part number 07272. Where operation of the Model T360U diverges from that of the Model T360, this addendum takes precedence. Please note that the T360 operation manual contains important SAFETY messages and the WARRANTY information for this instrument.
This page intentionally left blank ii
TABLE OF CONTENTS About This Addendum ................................................................................................................ i Revision History.......................................................................................................................... i Table of Contents ......................................................................................................................iii 1. Overview .................................................................
This page intentionally left blank.
1. OVERVIEW The Model T360U is a close derivative of the Model T360 CO Analyzer. The main differences are as follows: The optical bench is longer (14 m), making the instrument more sensitive at low CO2 levels. The instrument has a user-selectable full scale range of 0-100 PPB to 0-100 PPM of CO2. The rear panel has an extra pneumatic port for the exhaust of the purge gas to the GFC wheel.
PARAMETER AC Power SPECIFICATION 100 – 120V 50/60 Hz (120W) 220 – 240 V 50/60 Hz (144W) Environmental Conditions Installation Category (Over voltage Category) II Pollution Degree 2 Standard I/O 1 2 2 8 6 4 Optional I/O 1 USB com port 1 RS485 8 analog inputs (0-10V, 12-bit) 4 digital alarm outputs Multidrop RS232 3 4-20mA current outputs Analog Output Resolution 1 part in 4096 of selected full-scale voltage Certifications CE: IEC 61010-1:2001, EN61326 - Class A 1 2 3 4 5 Ethernet: 10/100Base-T R
4. PNEUMATIC CONNECTIONS This section provides information on the pneumatic configurations. Figure 4-1 shows the basic pneumatic flow of the T360U. Figure 4-1.
4.1. BASIC AND ZERO/SPAN VALVE OPTION CONFIGURATIONS CAUTION Sample and calibration gases should only come into contact with PTFE (Teflon), FEP, glass, stainless steel or brass. Do NOT operate without first removing dust plugs from rear panel fixtures. NOTE To prevent dust from entering the gas flow channels, your analyzer was shipped with small plugs inserted into each of the pneumatic fittings on the back panel. Remove these dust plugs and store for future use before proceeding.
Sample gas Pressure regulator, set output to 5-10 psig Needle valve, set flow to 800 cc/min (+/- 10%) Sample Chassis Exhaust Flowmeter Vent Span Span gas CO2 free gas Needle valve, set flow to 800 cc/min (+/10%) Pressure Regulator (5-10 psig output) Needle valve, set flow to 800 cc/min (+/10%) Pressure Regular (5-10 psig output) Pressure Span Zero To Purge From Purge 10-30 psig CO2 free gas Figure 4-3.
4.2. MAKING THE PNEUMATIC CONNECTIONS CAUTION Venting should be outside the shelter or immediate area surrounding the instrument. 1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long. 2. Attach sources of zero air and span gas 3. Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range.
5. RELAY ALARM OUTPUTS (STANDARD CONFIGURATION) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2 exceeded. AL4 is not used The relay alarm output AL1 is enabled all the time, whereas the AL2 and AL3 can be enabled/disabled by going into the diagnostics menu (with 929 password), then factory options, then turning conc. Alarm Relays ON or OFF.
This page intentionally left blank.
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 PARTNUMBER 037600000 037860000 039060100 039260501 040010000 040030300 040370000 041350000 042680000 042690000 042890100 043250100 043250300 043250400 045930000 050110000 052560000 052830200 055010000 058021100 066970000 067240000 067300000 067300100 067300200 067900000 068810000 069500000 072150000 072850000 074610100 074620100 077480300 037250000 036110400 033560000 033520000 032960000 019340600 016910000 016300600 016290000 010800000 010790000 0098
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 009450300 006110200 003291500 003290500 001761300 001760400 000941000 000940600 077490300 CN0000073 CN0000458 CN0000520 FL0000001 HW0000020 HW0000036 HW0000090 HW0000101 HW0000379 HW0000453 HW0000685 KIT000219 OP0000009 OR0000001 OR0000002 OR0000034 OR0000039 OR0000041 OR0000047 OR0000077 OR0000088 OR0000094 OR0000096 OR0000097 PS0000011 PS0000024 PS0000025 PU0000022 SW0000025 SW0000026 VA0000014 WR0000008 14 ASSY, ZERO/SPAN VALVES, CO ASSY, MOTOR WH
Interconnect List, T360U Reference: 073640100A FROM Cable PN Signal Assembly PN J/P Pin 03995 CBL, MOTOR TO RELAY PCA GFC Drive - A Relay PCA 041350000 J6 1 GFC Drive - B Relay PCA 041350000 J6 2 Motor Return Relay PCA 041350000 J6 3 Chassis Gnd Relay PCA 041350000 J6 4 040230100CBL, I2C, RELAY BOARD TO MOTHERBOARD I2C Serial Clock Motherboard 058021100 P107 3 I2C Serial Data Motherboard 058021100 P107 5 I2C Reset Motherboard 058021100 P107 2 I2C Shield Motherboard 058021100 P107 6 04146 CBL, SYNC DEMOD DG
Interconnect List, T360U Reference: 073640100A FROM Cable PN Signal Assembly PN J/P 06746 CBL, MB TO 06154 CPU GND Motherboard 058021100 P12 RX0 Motherboard 058021100 P12 RTS0 Motherboard 058021100 P12 TX0 Motherboard 058021100 P12 CTS0 Motherboard 058021100 P12 RS-GND0 Motherboard 058021100 P12 RTS1 Motherboard 058021100 P12 CTS1/485Motherboard 058021100 P12 RX1 Motherboard 058021100 P12 TX1/485+ Motherboard 058021100 P12 RS-GND1 Motherboard 058021100 P12 RX1 Motherboard 058021100 P12 TX1/485+ Motherboard
Interconnect List, T360U Reference: 073640100A FROM Cable PN Signal Assembly PN 06814 CBL ASSY, SYNC DEMOD , DC POWER DGND Relay PCA 041350000 +5V Relay PCA 041350000 AGND Relay PCA 041350000 +15V Relay PCA 041350000 AGND Relay PCA 041350000 -15V Relay PCA 041350000 +12V ret Relay PCA 041350000 +12V Relay PCA 041350000 06815 CBL ASSY, AC POWER AC Line Power Entry CN0000073 AC Neutral Power Entry CN0000073 Power Grnd Power Entry CN0000073 Power Grnd Power Entry CN0000073 AC Line Switched Power Switch SW0000
This page intentionally left blank.
07273A DCN6707 19