Portable Oxygen Monitor OPERATING & SERVICE INSTRUCTIONS MX300-I

Portable Oxygen Monitor Operation
Teledyne Analytical Instruments 15
Operation
Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR
DAMAGE. Check the unit and all included accessories for
broken or loose parts. If damaged, DO NOT USE. Notify
the shipper, and consult Teledyne Analytical Instruments.
Note: This equipment is internally powered using 3 AA batteries.
CAUTION: THE MX300-I, OXYGEN SENSOR AND ASSOCIATED
HARDWARE ARE NON-STERILE DEVICES. DO NOT
AUTOCLAVE THE INSTRUMENT OR SENSOR, AS
THIS WILL DAMAGE THE EQUIPMENT.
2.1 Setup
The MX300-I Portable Oxygen Monitor is suitable for use in many
medical applications. The unit is equipped with a stand and can be used
on a tabletop or wall mounted using mounting bracket. An optional pole
mount clamp is also available.
To set up and use your MX300-I monitor:
1. Install the sensor.
2. Install the batteries.
3. Calibrate the unit.
4. Set the alarms.
The control keys are designed for easy operation. A LOCK/UNLOCK
key has been supplied to prevent accidental changes to critical settings. This
eliminates unwanted changes in calibration or alarm settings from accidental
touching or bumping of the keys. To further reduce the possibly of incorrect
adjustments at least two keys must be pressed in order to modify a critical
calibration or alarm set point value.
Note: The ALARM SILENCE () and BATT TEST key continue
to operate normally when the lock feature is activated.