Portable Oxygen Monitor OPERATING & SERVICE INSTRUCTIONS MX300-I

Operation MX300-I
Teledyne Analytical Instruments 16
2.1.1 Sensor Installation or Replacement
Note: The R17MED oxygen sensor must be installed before the
oxygen analyzer/monitor can be operated
1. Remove the new sensor from its protective bag. Inspect the
sensor for damage or electrolyte leakage. If the sensor is
damaged, obtain a replacement. Do not use the defective
sensor as it may damage the unit.
WARNING: THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT
LET IT COME IN CONTACT WITH SKIN. IF IT DOES,
FLUSH AFFECTED AREA WITH WATER. DO NOT
ATTEMPT TO OPEN OR REPAIR THE SENSOR.
WARNING: THE SENSOR ALSO CONTAINS LEAD. LEAKING OR
EXHAUSTED SENSORS SHOULD BE HANDLED AND
DISPOSED OF IN ACCORDANCE WITH LOCAL
REGULATIONS. CONTACT TELEDYNE FOR THE
MATERIAL SAFETY DATA SHEET
2. Plug one end of the coiled cable into the jack receptacle on the
back end of the R17MED sensor and secure in place with the
capture nut located at the base of the connector. See Figure 2-1.
Figure 2-1: Installing the R17MED Sensor
3. Plug the other end of the coiled cable into the receptacle located on
the right side of the unit and secure it in place using the capture nut.
See Figure 2-2.