1.7 Operation Verification
The Model 401 Calibrator is now ready for operation.
1. Read Section 1.3 and all of Section 4 of the manual to understand the Calibrator operation.
2. Turn on the power by pressing the on/off switch on the front panel (see Figure 2-1). The
display should turn on and green (sample) status LED should be energized. The green LED
should blink indicating the instrument has entered the HOLD-OFF mode. The standby mode
can be entered immediately by pressing the EXIT button on the front panel. The red "fault"
light will also be on until the flows, temperatures and voltages are within operating limits.
Clear the fault messages.
3. After a 60 minute warm-up, review the TEST function values in the front panel display by
pushing the leftmost keyboard button labeled TEST. Not every TEST function is a diagnostic
of correct analyzer operation, therefore TEST functions not covered below can be ignored for
now.
4. O3 REF, O3 MEAS: TEST function values should be between 4200 mv and 4700 mv.
5. Pressure: 29 to 30 Inches-Mercury-Absolute at sea level. Other values will be displayed
depending on altitude of analyzer.
6. Sample Flow: 800 cc/min ± 10 %.
7. Sample Temp: Ambient temperature +10°
C.
8. Analytical Lamp Temp: 52° C ± 0.1 The computer drives the temp to this setpoint -it should
be this value and not vary.
9. O
3
GEN Lamp Temp: 48° C ± 1° C.
10. Box Temp: Ambient ± 10° C.
11. DC Power Supply: 2500 mV ± 100 mv. This is a composite of all of the DC voltages in the
instrument. The value is not important but it should be within the range indicated, and should
be constant.
12. If the TEST functions are within the limits given above the instrument should function
correctly. If there is a problem please read the manual and check your set-up. The Model 401
is now ready for calibration, see Section 3.
1-20