MODEL 319R OXYGEN DETECTOR INSTRUCTION MANUAL P/N M59608 ECO# 95-416 7/10/95 TELEDYNE BROWN EGINEERING Analytical Instruments
Copyright © 1993 Teledyne Brown Engineering Analytical Instruments All Rights Reserved.
Table of Contents 1.0 Introduction Main Features ............................................................................................... 1 Micro-Fuel Cell ............................................................................................ 2 Z-2 Cell Warranty Conditions ...................................................................... 2 2.0 Installation & Operation Precautions....................................................................................................
TELEDYNE BROWN EGINEERING Analytical Instruments
319R Introduction 1.0 Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures. Modular and compact design provide outstanding performance and reliable on-line monitoring and make the 319R virtually maintenance-free. A special range on the 319R enables you to calibrate in ambient air.
1.0 Introduction 319R 4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months. 5. Analog Voltage Output. The Model 319R provides customer connections to external devices, such as computers, recorders, alarms, etc., at 0–10 VDC, negative ground, plus contact closures for range identification. Custom voltage outputs, either grounded or isolated, are available as options. 6. Analog Current Output.
319R Installation & Operation 2.0 Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage. Notify the shipper if any damage is found. Check the unit, probe, and included accessories for broken or loose parts. 3. The sample must be free of entrained solids and water. A high humidity sample is ideal since it will prevent water loss from the cell electrolyte. 4.
2.0 Installation & Operation The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level of electromagnetic noise in the operating environment. Analysis Unit Connection The receptacle for the analysis unit cable is located in the lower lefthand corner of the rear panel. The 6-pin Minifit fitting is keyed to fit only one way in the receptacle; do not force it in.
319R Installation & Operation 2.0 Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuit connection. The explanation for the relay terminals abbreviations is as follows: NC = Normally closed C = Common NO= Normally open The alarm relay circuits are designed for failsafe operation, meaning that the relays are activated (energized) during normal operation.
2.0 Installation & Operation Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale. Thus, an alarm set to indicate 192 on the display while on the 200 PPM O2 range will activate at 192 PPM O2 on the 200 PPM O2 range and at 1920 O2 on the 2000 O2 range (96% of full scale). To set the alarms: 1. Place the range switch located below the front panel display in the 2000 PPM range position (or highest range available). 2.
319R Installation & Operation 2.0 Calibration To calibrate the 319R: 1. Place the power switch in the ON position. 2. Allow the LCD reading to stabilize. 3. Expose the sensor to ambient air (20.9%) or instrument grade air. If the analyzer goes overrange, the display will go blank and the LED to the left of the display will blink. Hold down the SPAN button until the LED stops blinking. 4. While holding down the SPAN AT 20.
2.
319R Maintenance 3.0 Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using the following procedure: 1. Disconnect the AC power and place the power switch located on the rear panel in the OFF position. 2. The fuse is located next to the power cord receptacle in the upper left-hand corner of the rear panel. Remove the power cord from the receptacle. 3.
3.0 Maintenance 319R 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing the knob marked FUSE. The fuse will be located inside the knob.
319R Maintenance 3.0 Warranty Conditions The Class Z-2 cell used in the 319R is warranted for six (6) months of service from the date of shipment. Customers having warranty claims must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty material or workmanship, the cell will be replaced free of charge. CAUTION: Evidence of tampering or abuse will render the warranty null and void.
3.
319R Appendix Appendix Specifications Ranges: Sensitivity: Signal Output: Display: Alarms: Accuracy: Response: Operating Temp. Range: Power: Enclosure: Cell Type: Cell Warranty: Cell Life (Expected): Dimensions: 0–200 PPM, 0–1999 PPM O2, and 0– 25% calibration range 1% of full scale Standard: 0–10 VDC, negative ground 4-20 mA, negative ground plus contact closures for range identification. Digital 7 segment, 3½-digit LCD with over range indicator. One high alarm, adjustable & fail-safe.
Appendix 319R Spare Parts List QTY. 1 1 1 4 4 1 P/N C57304 A33748 C6689-Z2 F1189 F51 A59264 DESCRIPTION PC board, main Thermistor assembly Micro-Fuel Cell, Z-2 Fuse, 1A (AC) Fuse, ½A (DC) Probe to analyzer cable, 6 feet A minimum charge of US $20.00 is applicable to spare parts orders. IMPORTANT: Orders for replacement parts should include the part number and the model and serial number of the system for which the parts are intended.
319R Appendix Drawing List C-57490 A-58154 D-57302 C-57305 C-57307 B-58123 Note: Outline diagram Probe interconnection diagram Electronic PCB schematic diagram Power supply PCB schematic diagram Power supply PCB assembly diagram Analysis unit (probe) outline 319R analyzers with 25% and 100% range modifications incorporate the following changes: 1. 2. 3. 4. 100% range modified circuit changes: JP2 - connects pins 1 and 2 R52 is 12.5K R30 is 7.5K R35 is 1.5K 1. 2. 3. 4.
Appendix 319R Material Safety Data Sheet Section I – Product Identification Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes.
319R Appendix Section IV – Fire and Explosion Hazard Data Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fire conditions. No specific agents recommended. Special Fire Fighting Equipment: Wear NIOSH/OSHA approved self-contained breathing apparatus and protective clothing to prevent contact with skin and eyes. Unusual Fire and Explosion Hazards: Emits toxic fumes under fire conditions.
Appendix 319R Signs and Symptoms of Exposure: Contact of electrolyte with skin or eyes will cause a burning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss of sleep, loss of appetite, metallic taste and fatigue. Carcinogenicity: Lead is classified by the IARC as a class 2B carcinogen (possibly carcinogenic to humans) OSHA: Where airborne lead exposures exceed the OSHA action level, refer to OSHA Lead Standard 1910.1025.
319R Appendix Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: Apron, face shield Ventilation: na Section IX – Disposal Both lead and potassium hydroxide are considered poisonous substances and are regulated under TSCA and SARA Title III. EPA Waste Number: D008 California Waste Number: 181 DOT Information: RQ Hazardous Waste Solid N.O.S. (Lead), 9, UN3077, PG III Follow all Federal, State and Local regulations.