6.SETUPMODE
The SETUP mode is mode is accessed by pressing the SETUP key momentarily so the words ‘Setup
Mode’ appear at the bottom of the display. The transmitter will return to the RUN mode when the
SETUPkeyispressedagainor60secondsafterthelastkeypress.
6.1FUNCTIONSUMMARYTABLE
Whenthetransmitterisin theSETUP modethe SETUPlight willbeonandthewords“SetupMenu”
willbeshownatthebottomofthedisplay.
ThefollowingtableshowstheSETUPfunctions:
Menu# Functionname Range Defaultvalue
1 Probe1Offset ‐6.0to+6.0mV 0.0mV
2 Probe2Offset ‐6.0to+6.0mV 0.0mV
3 LowerLineItems SeeChapter6.3.2
4 OxygenDisplayUnits Oxygen%/Oxygenppm Oxygen%
5 DampingFactor NoDampingto10Samples 5Samples
6 ProcessAlarms Enabled/Disabled Disabled
6.2CHANGINGOPTIONS
The purpose of having a menu driven SETUP mode is to allow for configuration of the transmitter
usingthegraphicaldisplayandkeypad.
OnceanoptionischangedandenteredusingtheENTERkeythatvalueimmediatelybecomesactive.
It is also stored into the non‐volatile memory within the BBRAM and will be loaded again at device
start‐up.Toresetthetransmitterconfigurationdefaultsseechapter10.2,ColdStart.
TochangeanoptionintheSETUPmenusystem:
1. SelecttheSETUPmodebypressingtheSETUP/RUNkeyonce.TheSETUPlightwillcomeonand
thedisplaywillhaveth
eformatshownbelow.Theoperationsofthekeysarenowtheoperations
writteninWHITEonthekeypad.Themenunameiswrittenatthebottomofthedisplay.
2. When the SETUP mode has been selected the required function can be found by using the
FUNCTIONUPandFUNCTIONDOWNkeys.
The options available for that function can be seen by using the OPTION UP & OPTION DOWN
keys.
3. WhentherequiredoptionisonthedisplaypresstheENTERkeytosavethatvalue.
Whenfinished,presstheSETUP/RUNkeytoreturntotheRUNmode.