ImportantNoticeRegarding9060 ProbeOption‐FIL3
WARNING:TheonlyidentifiablestandardforflamearrestersforgeneraluseisBritishStandardBS
EN12874:2001.BritishStandardBSEN12874:2001referstoanoperatingenvironmentupto150
DegreesCentigrade.
The FIL3 device optionally fitted to 9060 Heated Zirconia Probes (the “Probes" or "Probe")
operateinanenvironmentconsiderablygreaterthan150DegreesCentigrade.
Therefore, we know of no Australian, British, European or USA standard applicable to flame
arrestersortheirtestingabove150degreesCentigrade.Consequently,theFIL3devicecannotbe
certifiedasasafetydevice.
Theprobeisonlyonprobesduringthestartupprocessesofacombustionappliance.
TheTeledyneAnalyticalInstrumentsBurnerInterlockRelayfacility,whichisastandardpartofthe
Teledyne Analytical Instruments transmitter, is designed to be wired to the main safety shutoff
fuelvalvesinawaythatcanshutdowntheprobeheaterwhenthefuelvalvesareclosed.
TheriskofignitionofflammablegasmixtureatthehotendoftheProbecanonlybeminimizedby
correct use, maintenance and operation of the FIL3 device. The user of the FIL3 device is
responsibleforverificationandmaintenanceandcorrectuseandoperationoftheFIL3device.
THE USER AGREESTHAT IT USES THE PROBE AND THE FIL3 DEVICE AT ITS SOLE RISK. TELEDYNE
ANALYTICALINSTRUMENTSTOTHEFULLEXTENTPERMITTEDBYLAW,GIVESNOWARRANTIESOR
ASSURANCESANDEXCLUDESALLLIABILITY(INCLUDINGLIABILITYFORNEGLIGENCE)INRELATION
TOTHEPROBEANDTHEFIL3DE
VICE.
The user must ensure that it correctly follows all instructions in relation to the Probe and FIL3
device,correctlyunderstandsthespecificationsoftheProbeandFIL3deviceandensuresthatthe
ProbeandFIL3deviceareregularlyinspectedandmaintained.
FIL3equippedProbessho
uldbeinspectedatleastonceayearforcorrosionandmorefrequently
ifthereisanyreasontosuspectthatcorrosionmayhaveoccurred.