Photometric Analyzer OPERATING INSTRUCTIONS Model 5000B Photometric Analyzer DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM. HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING.
Model 5000B Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 917491580.
Photometric Analyzer Table of Contents Part I: Control Unit ................................ Part I: 1-1 Part II: Analysis Unit ............................ Part II: 4-1 Appendix .........................................................
Model 5000B iv Teledyne Analytical Instruments
Part I: Control Unit OPERATING INSTRUCTIONS Model 5000B Photometric Analyzer Part I: Control Unit NEMA 4 Bulkhead Mount Teledyne Analytical Instruments Part I: i
Model 5000B Photometric Analyzer Table of Contents 1 Introduction 1.1 Overview ........................................................................ 1-1 1.2 Typical Applications ....................................................... 1-1 1.3 Main Features of the Analyzer ....................................... 1-5 1.4 Operator Interface .......................................................... 1-6 1.4.1 Up/Down Switch .................................................... 1-6 1.4.2 Escape/Enter Switch .
Part I: Control Unit 3.3.12 Model ..................................................................... 3-15 3.3.13 Show Negative ...................................................... 3-15 3.4 The Zero and Span Functions ....................................... 3-16 3.4.1 Zero Cal ................................................................. 3-16 3.4.1.1 Auto Mode Zeroing ........................................ 3-16 3.4.1.2 Manual Mode Zeroing .................................... 3-17 3.4.1.
Model 5000B Photometric Analyzer iv: Part I Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit Introduction 1.1 Overview The Teledyne Analytical Instruments Model 5000B Control Unit, together with a 5000B Analysis Unit, is versatile microprocessor-based instrument. Part I, of this manual covers the Model 5000B General Purpose NEMA 4 Bulkhead Mount Control Unit. (The Analysis Unit is covered in Part II of this manual.) The Control Unit is for indoor/outdoor use in a nonhazardous environment only.
1 Introduction Model 5000B Ethanol Isopropanol Methanol Alkanes, including: 0–500 ppm Heptane Hexane Ammonia 0–1000 ppm and up Aromatics, including: 0–500 ppm Benzene Cumene Toluene Xylene Chlorinated Hydrocarbons including: 0–200 ppm Carbon Tetrachloride Ethyl Chloride Ethylene Dichloride Methyl Chloride Perchloroethylene Propylene Dichloride Trichloroethylene Vinyl Chloride Chloroprene 0–200 ppm Chloropicrin 0–200 ppm Deuterium Oxide 0–200 ppm Epichlorohydrin 0–2000 ppm Ethylene Glycol
Photometric Analyzer Part I: Control Unit Hydrogen Fluoride 0–10% Hydroperoxides 0–5% Kerosene 0–500 ppm Ketones 0–1000 ppm Methyl Acetate 0–1000 ppm Methyl Methacrylate 0–1000 ppm Oils 0–1% Olefins 0–500 ppm Pentane 0–300 ppm α-Picolene 0–300 ppm Phenol 0–1000 ppm Polyols 0–500 ppm Propylene Glycol 0–500 ppm Propylene Oxide 0–200 ppm Sulfinol 0–15% Sulfur Dioxide 0–1000 ppm Vinyl Acetate 0–2% NOTE:Range may be higher or lower per application.
1 Introduction Model 5000B Hydroxyl Value Hydrocarbons Hydrogen Chloride Hydrogen Fluoride Hydroxyl Value Ketones Olefins Oximes Epoxides Methylene Ketones 1-4: Part I Teledyne Analytical Instruments
Photometric Analyzer 1.3 Part I: Control Unit Main Features of the Analyzer The Model 5000B Photometric Analyzer is sophisticated yet simple to use. The main features of the analyzer include: • A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator. • High resolution, accurate readings of concentration from low ppm levels through to 100%. Large, bright, meter readout. • Versatile analysis over a wide range of applications.
1 Introduction 1.4 Model 5000B Operator Interface All controls and displays on the standard 5000B are accessible from outside the housing. The instrument has two simple operator controls. The operator has constant feedback from the instrument through an alphanumeric display, and a digital LED meter. The displays and controls are described briefly here and in greater detail in chapter 3. See Figure 1-1. 1.4.
Photometric Analyzer -Range Part I: Control Unit This function selects whether analyzer is autoranging or locked on one range. -Standby Places the analyzer in a sleep mode. WARNING: The power cable must be disconnected to fully remove power from the instrument. Figure 1-1: Model 5000B Controls, Indicators, and Connectors Digital Meter Display: The meter display is a Light Emitting Diode LED device that produces large, bright, 7-segment numbers that are legible in any lighting.
1 Introduction Model 5000B Alphanumeric Interface Screen: The backlit VFD screen is an easyto-use interface between operator and analyzer. It displays values, options, and messages for immediate feedback to the operator. 1.5 Control Section Interface Panel The Control Section interface panel, shown in Figure 1-2, contains the electrical terminal blocks for external inputs and outputs. The input/output functions are described briefly here and in detail in the Installation chapter of this manual.
Photometric Analyzer Part I: Control Unit Figure 1-2: Model 5000B Interface Panel Teledyne Analytical Instruments Part I: 1-9
1 Introduction 1-10: Part I Model 5000B Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit Installation Installation of Model 5000B Analyzers includes: 1. Unpacking, mounting, and interconnecting the Control/Analysis Section 2. Making gas connections to the system 3. Making electrical connections to the system 4. Testing the system. This chapter covers installation of the Control Section. (Installation of the Analysis Section is covered in Part II of this manual.) 2.
1 Introduction Model 5000B Figure 2-3: Interface Panel of the Model 6600 Control Section For safe connections, ensure that no uninsulated wire extends outside of the terminal blocks. Stripped wire ends must insert completely into terminal blocks. No uninsulated wiring should come in contact with fingers, tools or clothing during normal operation. Primary Input Power: The power supply in the Model 6600 will accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed connections.
Photometric Analyzer Part I: Control Unit Figure 2-4: Primary Input Power Connections 115VAC, Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T type (slow blow) fuses. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation (See Fuse Replacement in chapter 4, maintenance.) Analog Outputs: There are eight DC output signal connectors on the ANALOG OUTPUTS terminal block. There are two connectors per output with the polarity noted. See Figure 2-5.
1 Introduction Model 5000B 4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA = Range 3. Figure 2-5: Analog Output Connections Examples: The analog output signal has a voltage which depends on the sample concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.
Photometric Analyzer Part I: Control Unit 3 0.3 8.8 4 0.4 10.4 5 0.5 12.0 6 0.6 13.6 7 0.7 15.2 8 0.8 16.8 9 0.9 18.4 10 1.0 20.0 To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current when using the current outputs) to represent a particular range. Table 2-2 gives the range ID output for each analysis range.
1 Introduction Model 5000B Normally closed Normally open Moving contact Normally open Moving contact Figure 2-6: Types of Relay Contacts The connectors are: Threshold Alarm 1: Threshold Alarm 2: System Alarm: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below thresh old). • Can be configured as fail-safe or non-fail-safe. • Can be configured as latching or nonlatching. • Can be configured out (defeated).
Photometric Analyzer Part I: Control Unit accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See Remote Calibration Protocol below.) Zero: Floating input. 5 to 24 V input across the + and – terminals puts the analyzer into the ZERO mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows ZERO mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately.
1 Introduction Model 5000B When CRC opens again, zero and span are done, and the sample is being analyzed. Note: The Remote Bench terminal strip (section 3.6 Part III) provides signals to ensure that the zero and span gas valves will be controlled synchronously. Range ID Relays: Four dedicated RANGE ID CONTACT relays .
Photometric Analyzer • Part I: Control Unit Which alarms - if any - are tripped (AL-x ON) Each status output is followed by a carriage return and line feed. Three input functions using RS-232 have been implemented to date. They are described in Table 2-4. Table 2-4: Commands via RS-232 Input Command Description as Immediately starts an autospan. az Immediately starts an autozero. st Toggling input.
1 Introduction • Model 5000B Check that sample pressure typically between 0 and 30 psig, according to the requirements of your process. Power up the system, and test it by performing the following operation: 1. Repeat the Self-Diagnostic Test. 2. Zero the instrument. 3. Span the instrument. Consult calibration recommendation in the addendum for your particular application whether liquid or gas phase monitoring. For steps 2 and 3, refer to part II for calibration.
Photometric Analyzer Operation /Control Unit 3 Operation 3.1 Introduction Although the Model 5000B is usually programmed to your application at the factory, it can be further configured at the operator level, or even, cautiously, reprogrammed. Depending on the specifics of the application, this might include all or a subset of the following procedures: • Setting system parameters: • Establish a security password, if desired, requiring Operator to log in (secure in safe file for referrence).
3 Operation/Control Unit Model 5000B The Enter button is used to accept any new entries on the VFD screen. The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button. Figure 3-1 shows the hierarchy of functions available to the operator via the function buttons. The six functions of the buttons on the analyzer are: • Analyze. This is the normal operating mode.
Photometric Analyzer Operation /Control Unit 3 System Dig_filt Set Digital Filter SELF-TEST Self-Testin Progress PWD Enter Password LOGOUT SecureSystem setup not allowed Self-Test Results Change Yes/No Change Password Verify Password Enter MORE AUTOCAL FILSOL Span/Zerostatus and <>setup Span/Zero SolenoidorFilter TRACK or HOLD Set track or hold output CAL-HOLD TIMER Set cal.
3 Operation/Control Unit Model 5000B the appropriate point in the procedure, in a Monospaced type style. Push-button names are printed in Oblique type. 3.3 The System Function The subfuctions of the System function are described below.
Photometric Analyzer Operation /Control Unit 3 Note: Before setting up an AUTO-CAL, be sure you understand the Zero and Span functions as described in section 4.4, and follow the precautions given there. Note: If you require highly accurate AUTO-CAL timing, use external AUTO-CAL control where possible. The internal clock in the Model 5000BF is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day. To setup an Auto–Cal cycle: Choose System from the Function buttons.
3 Operation/Control Unit 3.3.2 Model 5000B Password Protection Before a unique password is assigned, the system assigns TAI by default. This password will be displayed automatically. The operator just presses the Enter key to be allowed total access to the instrument’s features. If a password is assigned, then setting the following system parameters can be done only after the password is entered: alarm setpoints, assigning a new password, range/application selections, and curve algorithm linearization.
Photometric Analyzer Operation /Control Unit 3 arrow keys to change the letters to the proper password. Press Enter to enter the password. In a few seconds, you will be given the opportunity to change this password or keep it and go on. Change Password? =Yes =No Press Escape to move on, or proceed as in Changing the Password, below. 3.3.2.2 Installing or Changing the Password If you want to install a password, or change an existing password, proceed as above in Entering the Password.
3 Operation/Control Unit Model 5000B When you have finished typing the new password, press Enter. A verification screen appears. The screen will prompt you to retype your password for verification. Enter PWD To Verify: AAA Use the arrow keys to retype your password and press Enter when finished. Your password will be stored in the microprocessor and the system will immediately switch to the Analyze screen, and you now have access to all instrument functions.
Photometric Analyzer 3.3.4 Operation /Control Unit 3 System Self-Diagnostic Test The Model 5000BF has a built-in self-diagnostic testing routine. Preprogrammed signals are sent through the power supply, output board, preamp board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 1024. (See System Self Diagnostic Test in chapter 4 for number code.
3 Operation/Control Unit Model 5000B 3.3.5 The Model Screen Move the < > arrow key to MORE and press Enter. With MODEL blinking, press Enter. The screen displays the manufacturer, model, and software version information. 3.3.6 Checking Linearity with ALGORITHM From the System Function screen, select ALGORITHM, and press Enter. sel rng to set algo: —> Ø1 Ø2 Ø3 <— Use the < > keys to select the range: 01, 02, or 03. Then press Enter. Fluid Use: SO2 Range: Ø — 10% Press Enter again.
Photometric Analyzer Operation /Control Unit 3 The manual mode only requires entering the values for each correction point into the microprocessor via the front panel buttons. Again, the number of points required is determined by the user. 3.3.7 Digital Filter Setup The 5000BF has the option of decreasing or increasing the amount filtering on the signal. This feature enhances the basic filtering done by the analog circuits by setting the amount of digital filtering effected by the microprocessing.
3 Operation/Control Unit 6 9.0 7 14.0 8 25.0 9 46.0 10 90.0 Model 5000B At a setting of “zero”, the response time is purely set by the electronics to 4.5 seconds. The numbers above can and will change depending on application and they merely serve to illustrate the effect of the digital filter. 3.3.8 Filter or Solenoid Setup The 5000BF can be spanned or zeroed by calibration fluids or by optical filters. The proper calibration method should be set at the factory.
Photometric Analyzer 3.3.9 Operation /Control Unit 3 Hold/Track Setup The 5000BF has ability to disable the analog outputs and freeze the display while undergoing a scheduled or remote calibration. The 5000BF will track changes in the concentration if calibration is started through the front panel. To setup this feature, the operator must: 1. Press the System key to start the System function: DIG_FILT SELF-TEST PWD LOGOUT MORE 2. Using the Right or Left arrow keys, select MORE and press Enter.
3 Operation/Control Unit Model 5000B This function and the TRACK/HOLD feature will prevent false alarms while performing remote or autoscheduled calibrations. These functions are not applicable if the calibration is initiated through the front panel. To enter the Calibration/Hold Timer function, you must: 1. Press the System key to start the System function: DIG_FILT SELF-TEST PWD LOGOUT MORE 2.
Photometric Analyzer Operation /Control Unit 3 or AUTOCAL FILSOL HOLD CAL-HOLD-TIMER MORE 3. Using the Right or the Left arrow keys, select MORE and press Enter. The third System screen appears: ALGORITHM APPLICATION MODEL OUTPUT: 4 MA or ALGORITHM APPLICATION MODEL OUTPUT: 20 MA OUTPUT: 4 MA and OUTPUT: 20 MA can be toggled by moving on that field and pressing the Up or Down key. 4 mA output should be calibrated first and 20 mA output afterwards. 4.
3 Operation/Control Unit Model 5000B 3.3.13 Show Negative The analyzer defaults to not to show negative readings on the analyze mode only. This affects the analog outputs too by pressing the UP or DOWN key, the analyzer can be set to display negative readings, on the SHOW_NEG field of the system menu. 3.4 The Zero and Span Functions The Model 5000BF can have as many as three analysis ranges plus a special calibration range (Cal Range).
Photometric Analyzer Operation /Control Unit 3 zero calibration is to be performed automatically or manually. Use the DÑ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears, blinking, on the display. Select zero mode: AUTO Press Enter to begin zeroing. ####.## % Slope=#.### SO2 C—Zero The beginning zero level is shown in the upper left corner of the display.
3 Operation/Control Unit Model 5000B ####.## % SO2 Zero adj:2048 C—Zero The analyzer goes through C–Zero, F–Zero, and S–Zero. During C– Zero and F–Zero, use the DÑ keys to adjust displayed Zero adj: value as close as possible to zero. Then, press Enter. S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto Mode Zeroing, above. It calculates the differences between successive samplings and displays the rate of change as Slope= a value in parts per million per second (ppm/s). ####.
Photometric Analyzer 3.4.2 Operation /Control Unit 3 Span Cal The Span function on the menu is used to span calibrate the analyzer. Span calibration can be performed in either the automatic or manual mode. Make sure the span fluid is flowing to the instrument. 3.4.2.1 Auto Mode Spanning Observe all precautions in sections 3.4 and 3.4.2, above. Press Span to enter the span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually.
3 Operation/Control Unit Model 5000B Use the DÑ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen. Span Val: 2Ø.ØØ % To begin span Use the < > arrow keys to toggle between the span concentration value and the units field (%/ppm). Use the DÑ arrow keys change the value and/ or the units, as necessary. When you have set the concentration of the span fluid you are using, press Enter to begin the Span calibration.
Photometric Analyzer Operation /Control Unit 3 Select zero mode: AUTO or Select zero mode: MAN Select whether you want the instrument to do an automatic or manual zero. If you do an automatic zero, the instrument does the zero by itself. If you do a manual zero you must manually enter inputs to the instrument to accomplish the zero, see in the corresponding section of the manual on how these two functions differ. When the Enter key is pressed, the following menu will appear: Zero off: 0.
3 Operation/Control Unit Model 5000B offset to be entered. 4. Do a manual run to check. Reintroduce the zero calibration fluid. Start a zero on the analyser but this time enter the offset value. 5. At the end of the zero function, check that the instrument reads the entered offset. 6. Reintroduce the process background fluid mix to the 5000BF sample cell in the Analyse mode. It should read close to zero once the reading is stable (+/- 1% error of full scale).
Photometric Analyzer Operation /Control Unit 3 • Both high (high and high-high) alarms, or • One high and one low alarm, or • Both low (low and low-low) alarms. 2. Are either or both of the alarms to be configured as failsafe? In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode. 3.
3 Operation/Control Unit • • • • Model 5000B • Latching? Ltch:Y/N (Yes/No). To define the setpoint, use the < > arrow keys to move the blinking over to AL1: ####. Then use the DÑ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing. To set the other parameters use the < > arrow keys to move the blinking over to the desired parameter. Then use the DÑ arrow keys to change the parameter.
Photometric Analyzer Operation /Control Unit 3 Select range mode: MANUAL If above screen displays, use the DÑ arrow keys to Select MANUAL, and press Enter. Select range to run —> Ø1 Ø2 Ø3 CAL<— Use the < > keys to select the range: 01, 02, 03, or CAL. Then press Enter. Fluid use: SO2 Range: Ø — 10 % Use the < > keys to toggle between the Range: low-end field and the Range: high-end field. Use the DÑ keys to change the values of the fields.
3 Operation/Control Unit Model 5000B The concentration ranges can be redefined using the Range function Manual screen, and the application fluides can be redefined using the System function, if they are not already defined as necessary. CAUTION: Redefining applications or ranges might require relinearization and/or recalibration. To setup automatic ranging: Press Range key to start the Range function.
Photometric Analyzer Operation /Control Unit 3 Figure 3-2 illustrates these schemes graphically. 0 0.01 0.
3 Operation/Control Unit 3.7 Model 5000B The Analyze Function Normally, all of the functions automatically switch back to the Analyze function when they have completed their assigned operations. Pressing the Escape button in many cases also switches the analyzer back to the Analyze function. Alternatively, you can press the Analyze button at any time to return to analyzing your sample.
Photometric Analyzer Operation /Control Unit 3 AUTOCAL FILSOLL HOLD CAL-HOLD-TIMER MORE Select MORE and press ENTER one more time ALGORITHM MORE OUTPUT: APPLICATION 4MA Now you will be able to select the APPLICATION and ALGORITHM set-up functions. 3.8.1 The Set Range Screen The Set Range screen allows reprogramming of the three analysis ranges and the calibration range (background fluid, low end of range, high end of range, and % or ppm units).
3 Operation/Control Unit Model 5000B scale setting until amplifier saturation is reached. Below amplifier saturation, the overrange readings are accurate UNLESS the application uses linearization over the selected range. To program the ranges, you must first perform the four steps indicated at the beginning of section 3.8 Programming. You will then be in the second System menu screen. ALGORITHM APPLICATION MORE OUTPUT: 4MA Use the < > arrow keys again to move the blinking to APPLICATION and press Enter.
Photometric Analyzer Operation /Control Unit 3 to the analyzer from the computer. 3.8.2 The Curve Algorithm Screen The Curve Algorithm is a linearization method. It provides from 1 to 9 intermediate points between the ZERO and SPAN values, which can be normalized during calibration, to ensure a straight-line input/output transfer function through the analyzer. Each range is linearized individually, as necessary, since each range will usually have a totally different linearization requirement.
3 Operation/Control Unit Model 5000B Repeat the above procedure for each of the data points you are setting (up to nine points: 0-8). Set the points in unit increments. Do not skip numbers. The linearizer will automatically adjust for the number of points entered. When you are done, Press ESCAPE. The message, Completed. Wait for calculation, appears briefly, and then the main System screen returns. To end the session, send: st st to the analyzer from the computer. 3.8.2.
Part I: Control Unit/Analysis Unit Maintenance 3 Maintenance Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing filter elements and fuses, and recalibration. WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL. 4.1 Fuse Replacement The 5000B requires two 5 x 20 mm, 6.3 A, F type (Fast Blow) fuses. The fuses are located inside the main housing on the Electrical Connector Panel, as shown in Figure 4-3. To replace a fuse: 1.
3 Maintenance Model 5000B Photometric Analyzer Figure 4-1: Removing Fuse Block Cap and Fuse from Housing 2. Replace fuse by reversing process in step 1. 4.2 System Self Diagnostic Test NOTE: Always run self diagnostic with the intended zero fluid. 1. Press the System button to enter the system mode. 2. Use the < > arrow keys to move to More, and press Enter. 3. Use the < > arrow keys to move to Self-Test, and press Enter.
Part I: Control Unit/Analysis Unit Maintenance 3 Preamp 0 OK >0 means that some of the ten gains of the amplifier have a large offset. The number is a code that tells which of the ten gtains have the large offset. Cell (detector) 0 OK >0 means that the analyzer failed to balance the measuring and the reference signals, the larger the number the farther off they were.4.3 Major Internal Components The major components in the Control Unit are shown in Figure 4-3.
3 Maintenance Model 5000B Photometric Analyzer WARNING: HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
Part II: Analysis Unit OPERATING INSTRUCTIONS Model 5000B Photometric Analyzer Part II: Analysis Unit NEC or ATEX Type 5000B - GP, Rack, Panel (Integral or Remote) 5000 - GP, Bulkhead (Z-Purged in Div II or Zone 2 areas) I, II, B, C, D (Integral) 5000B-(X-Purged, 1,1, B, C, D) (Integral) 5000B-(Purged, Zone 1) (Integral) Teledyne Analytical Instruments Part II: i
Model 5000B Photometric Analyzer Table of Contents 4 Operational Theory 4.0 Operations ..................................................................... 4-1 4.1 Control Function ............................................................ 4-2 4.1.1 Analysis Section .................................................... 4-2 4.1.2 Explosion-Proof Version: Control Module .............. 4-2 4.2 Start Up.......................................................................... 4-3 4.2.
Part II: Analysis Unit Appendix A-1 Specifications ................................................................ A-1 A-2 Recommended 2-Year Spare Parts List ......................... A-4 A-3 Drawing List ...................................................................
Model 5000B Photometric Analyzer iv: Part II Teledyne Analytical Instruments
Operations/Analysis Unit 4 4.0 Operations Before shipment, TAI calibrates the analyzer for your application when feasible. Calibration data is listed in the Appendix. Prior to calibration, TAI checks the electronics of the analyzer and makes all of the necessary internal printed circuit board adjustments. Calibration is performed to determine the proper or close proximity zero and span settings, and also to verify that the analyzer response is linear.
4 Operations/Analysis Unit 4.1 Control Functions 4.1.1 Analysis Section, fully explosion proof, Z or X purged Nema enclosure or Cenelec Purged (pending). The control functions for the analysis section are located on the Front of the Enclosure module: 1. POWER ON/OFF: This function is hard wired into the enclosure for power to the control/analysis section. 2.
Operations/Analysis Unit 4 The control unit digital display must handle functions for both the analysis section and the control module. 4.2 Start-p Information contained in this paragraph is based on the premise that the analyzer has been properly installed and that it is in operable condition. If difficulties arise during start-up, it is probable that some form of damage has incurred during shipment or some installation error has inadvertently been made.
4 Operations/Analysis Unit 4.2.2 Pre-Start-up Electrical Checkout After the preliminary procedures have been accomplished (refer to Preliminary Inspection and Control Settings, above), the integrity of the system interconnection and the power sources must be verified before attempting the analytical start-up procedures. The observations and measurements described in the following paragraphs are vital to the operation of the analyzer.
Operations/Analysis Unit 4 3. If the analyzer section has been equipped with the automatic zero SOL option, the zero solenoid device in the accessory sampling system should have energized the instant power was established. The device (or devices) should be energized because the mode switch has been preset to the ZERO position. 4.
4 Operations/Analysis Unit 4.3 Calibration Standardization Fluids Two standardization fluids are necessary to calibrate the analyzer: 1. Zero Fluid: The zero standard fluid must have a composition similar to the sample, and ideally, contains none of the components of interest. The zero fluid should be laboratory analyzed to determine its composition. The exact composition must be known, as the accuracy of the analysis can be no better than the knowledge of the standardization fluid. 2.
Operations/Analysis Unit 4 chemical background has changed. In some cases, TAI is not able to duplicate the background your sample for purposes of optically balancing the analyzer before shipment. In these cases, it is necessary for you to screen for an approximate balance, and to then electronically adjust R3 for precise balance. To do so: 1. Reset R3 to its midpoint. 2. Re-screen the filter wheel, as necessary, to obtain a measuring voltage within 10% of the reference voltage.
4 Operations/Analysis Unit instrument is set up to analyze, then the SPAN control must be adjusted for a full scale reading, i. e., 200 ppm water in EDC. 10. Re-check the ZERO setting with zero fluid. 11. If desired, the linearity of the analyzer can be checked with a fluid intermediate in concentration between the zero and the span fluid. 12. The analyzer is now calibrated.
Operations/Analysis Unit 4 A zero may be established with a reproducible simulated (offset) zero absorbance fluid in the sample cell of the analyzer. When possible this fluid should be stable for H2O content and should approximate the process being measured to minimize the offset zeroing, otherwise, dry air or nitrogen can be used. The analyzer must first be calibrated for sensitivity full scale by using known certified/analyzed samples containing the component (analyte(s) to be measured.
4 Operations/Analysis Unit as the off-line method used to verify the process variables. Below is an example of an on-line calibration when measuring 0-200ppm water in liquid TDI (assumes TDI is a toxic liquid at standard temperature and pressure conditions). The first sample should be taken when the analyte (H20) concentration is known to be low (approximately O-.002 % water).
Operations/Analysis Unit 4 optical filters. These optical filters are stable and are chosen to indicate the measured component at the required instrument wavelengths. Therefore, future, faster, yet much simpler calibration zero and span checks can be made without the unnecessary tedious on-line calibration described above nor the consumption of expensive calibration fluids which may be difficult to handle or obtain, are unstable, or toxic in nature.
4 Operations/Analysis Unit period). In this case, nitrogen or dry air (typically-100OF dewpoint or 2ppm water) can be used. Manually switch in the zero offset check from the control unit to operate the zero solenoid flag (depending upon the application, this may or may not simultaneously insert a zero flag into the light path; indeed, 90% of the time it is not needed).
Operations/Analysis Unit 4 always be corrected using the calibration zero and/or span features. SUMMARY The zero and/or span calibration of 5000 series photometric Teledyne analyzers can be quickly verified manually by using Zero and/or calibration zero offsets and/or span flags. Refer to typical Teledyne Analytical 1475 sample system below, 3-way valves 1,2,+ 3 below are used to check internal calibration using zero and/or span filters with N2.
4 Operations/Analysis Unit or 85 ppm simulated water using span flag on top of dry liquid TDI delta 80 ppm agreement on slope sensitivity. or 0 ppm water in N2 5 ppm water on the 5000B meter or output with zero-flag/offset intro. 0 ppm water in N2 delta 80 ppm simulated sensitivity change of water. Note: the span flag checks the sensitivity change of the 5000B and not the zero drift of the analyzer.
Operations/Analysis Unit 4 1 Sample Return Sample Flowmeter .
4 Operations/Analysis Unit ATTACHMENT 5000B/5020 NIR ANALYZER Quote "Exceptions" and "LIQUID PHASE Conditions" for this application. 1 Response Time is proportional to sample system design for take-off distance, process pressure, line size, sequencing and bypass, by-pass analyzer flow design, dead-volumes/tee's, sample cell volume and instrument electronics. 2 Maximum cell operation pressure to be 175 psig, based upon standard Sapphire window cell design. Higher pressures available.
Operations/Analysis Unit 4 4.4 Operational Theory Cont'd The energy source for the analyzer is most commonly provided by as a high intensity quartz iodine lamp located in the source module. Quartz iodine was chosen because it produces sufficient NIR to operate the system and maintains a nearly constant brightness over its lifetime. (See Figures 2-1 and 2.2).
4 Operations/Analysis Unit direct meter indication, which greatly eases the task of balancing the system during initial system installation and periods of calibration. The DC voltage levels are fed to a logarithmic ratio amplifier which produces a voltage output that is proportional to the logarithm of the ratio of the two DC input voltages. This output voltage, directly proportional to the concentration of sample, is, within certain limits, a linear function of the concentration.
Operations/Analysis Unit 4 Figure 4-1. Optical System 4.4.2 Sample cell The sample cell, generally constructed of 316SS, is located in the path of the NIR radiation, between the source and the detector modules. Each compound in the sample path exhibits its own characteristic absorption spectrum. Cell spacer thicknesses will vary depending upon the absorbance of the component of interest at the measuring wavelength.
4 Operations/Analysis Unit 2. The entire sample module is separately controlled to maintain the sample temperature during analysis. 3. The sample is temperature compensated for by software while inputing the temperture value at the sample cell.
From Detector Module Operations/Analysis Unit 4 15 VDC MEAS REF ZERO Power S1 5A Coarse Zero Control To Log Selec tor Switc h To Meter Driver Amplifier 115 VAC to Sourc e Module 115 VAC to Detector Module Line Voltage Regulator Transformer 115 VAC 60 Hz Input Figure 4-3 Preheater Temp. Control Thermistor and Heater Sample Module Temp. Control Thermistor and Heater Detector Module Temp.
4 Operations/Analysis Unit 4.4.3 Power See Figure 4-3. The power to the analyzer unit, enables the switching functions, and temperature control for the detector compartment. With the microprocessor control unit, the following information is provided. Figure 2-3. Power Module - Block Diagram The temperature in the detector compartment is controlled by inputs from a RTD device which feeds to a PID controller capable of set points of temperature versus temperature readout. 4.4.
Part II: Maintenance/Analysis Unit Maintenance 5 Maintenance 5.0 Routine Maintenance 5.1 Automatic operation and routine operational duties The system operates continuously without adjustment. Under normal conditions, after you program the system for automatic operation, only routine maintenance procedures are necessary. The most common failure condition is a temporary interruption of the power serving the instrument.
5 Maintenance Model 5000B Photometric Analyzer 3. Verify that the chart recorder contains a normal display. 4. Verify that the recorder has a sufficient supply of chart paper and ink. 5.3 Routine Maintenance Keep the sample lines and components, including the measuring cell within the analyzer sample module, free of deposits and leaks.
Part II: Maintenance/Analysis Unit 5.5 Maintenance 5 Service Procedures and Adjustments 5.5.1 Electronics TAI aligns the system’s electronics. However, you may need to touch up the circuitry, using the following procedure. Equipment Required: Oscilloscope (dual trace is preferred, but not required) To observe oscilloscope test points switch the vertical input selector of the scope to DC. Switch to AC to observe the demodulator switch signals.
5 Maintenance Model 5000B Photometric Analyzer If you open the detector module, keep stray light out by covering the opening with a dense black cloth. If you do not take this precaution, the result is a misinterpretation of the scope patterns. On general-purpose systems, the scope test points are in the bottom of the detector module and are accessible without opening the module. 5.5.
Part II: Maintenance/Analysis Unit Maintenance 5 The procedure is purely mechanical and consists of adjusting the amount of light passing through either the measuring or reference filter, never both. Screens (wire mesh) of varying density are used for this operation and are part of the small took kit accompanying the instrument. 1. Observe the oscilloscope and judge if optical balancing is needed. When the difference is less than 1 volt, balancing is not required.
5 Maintenance Model 5000B Photometric Analyzer 7. After the peaks are balanced, adjust the gain control until the tallest of the two peaks is 8 to 9 volts. The peaks should still be within1 volt of each other. 8. It is always good practice to operate the analyzer with as low a gain as possible. Therefore, with the gain control just barely off its stop, once again remove or add screens in the light path to obtain as high a voltage as possible without exceeding 9 volts for the highest peak.
Part II: Maintenance/Analysis Unit Maintenance 5 5.5.8 Integrated Reference and Measuring Signals You can observe the reference and measuring signal at the first stage of integration by connecting the scope probe to TP6 (reference signal) and TP7 (measuring signal) at the detector unit. A dual trace scope is advantageous but not required for this observation. The test points’ significance is that they reveal proper switch action.
5 Maintenance Model 5000B Photometric Analyzer WARNING: DANGEROUS HIGH VOLTAGES ARE PRESENT AT THESE TERMINALS. TRAINED PERSONNEL MUST REMOVE THE SILKSCREEN COVER ONLY. EXERCISE EXTREME CAUTION. The first strip terminal has three contacts labeled N, G and H. The labels stand for Neutral, Ground, and Hot. This is the AC power strip terminal. It feeds AC power to other components of the Model 6000B System, such as the D2 lamp power supply, heater, and temperature controller PCB.
Part II: Maintenance/Analysis Unit Maintenance 5 +15 VDC: Power Supply voltage fed to the photodetector preamplifier, +15 VDC. -15 VDC: Power Supply voltage fed to the photodetector preamplifier, -15 VDC. COM: Common reference to the +/- 15 VDC and the -230 VDC power supplies. SPAN FLTR: Span filter signal, AC voltage. SPAN SOL: Span solenoid signal, AC voltage. ZERO FLTR: Zero filter signal, AC voltage. ZERO SOL: Zero solenoid signal, AC voltage.
5 Maintenance 5-10: Part II Model 5000B Photometric Analyzer Teledyne Analytical Instruments
Photometric Analyzer, 5000B Appendix Appendix A-1 Specifications 5000B Digital Control Module: Ranges: Four Programmable Ranges, field selectable within limits (application dependent) and Auto Ranging Display: 2 line by 20 alphanumeric VFD accompanied by 5 digit LED display Signal Output: Two 0-1V DC (concentration and range ID) Two 4-20mADC isolated (concentration and range ID) RS232 Alarm: Two fully programmable concentration alarm set points and corresponding Form C, 3 amp contacts.
Appendix Models 5000B Typical Analytical Performance Specifications: (will vary per application) Accuracy: ±2% of full scale or better, application dependent Noise: Less than ±1% Diurnal: Less than 1% per 20oF (10oC (source/detector dependent) Sample Cell: Stainless steel with Sapphire window standard. Other materials available. Cell Length: .01 to 10 inches Flow Rate: 50 to 1500 cc/min Light Source: TungstenMini-Lamp, optional Quartz Halogen Sensitivity: .015 to 3 absorbance units.
Photometric Analyzer, 5000B Appendix Enclosure Purging: Purge air startup*:40 – 80 PSI (3 – 6 BAR), 0.
Appendix Models 5000B A-2 Recommended 2-Year Spare Parts List QtyP/NDescription C-75825A Motherboard, Control Unit C-67990 Amplifier, Control Unit D-67990 5000B Interface PCB 1 A-9306 Differential Power Supply 1 C-40265A Measuring PCB 1 L323 UV LAMP, 5000B 1 B77609 Parabolic Reflector Assy.
Photometric Analyzer, 5000B Appendix 16830 Chestnut Street City of Industry, CA 91749-1580 Phone (626) 934-1500, Fax (626) 961-2538 TWX (910) 584-1887 TDYANYL COID Web: www.teledyne-ai.com or your local representative. A-3 Drawing List, 5000B SEE MANUAL ADDENDUM OF Q FOR COMPLETE DRAWING LIST.
Appendix A-6 Models 5000B Teledyne Analytical Instruments