INSTRUCTION MANUAL PORTABLE OXYGEN MONITOR TED 191

15
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
S
ECTION
T
HREE
C
LEANING AND
S
TERILISATION
3.1 C
LEANING THE
TED 191 P
ORTABLE
O
XYGEN
M
ONITOR
The instrument itself should be cleaned using a damp cloth. In the event of an excessive
build-up of dirt, a mild alcohol solution (isopropyl) may be used to wipe over the instrument,
which should be allowed to air-dry afterwards. In extreme cases, the monitor casing may be
cleaned using a mild detergent.
Note: The Accurox Oxygen monitor cannot be liquid sterilised or autoclaved –
irreparable damage may result.
3.2 S
TERILISING THE
T-7 O
XYGEN
S
ENSOR AND
C
ONNECTING
C
ABLE
The T-7 oxygen sensors and connecting cables used on the instrument may be gas
sterilised using low temperature ethylene oxide. A vacuum must not be drawn on the
sensor during the sterilisation process.