End Threshold Conguration
This threshold is the end threshold of optical ber. If the end threshold equals 3.0 dB, then the rst event with insertion loss ≥ 3
dB will be considered the end of the optical ber. If the value is set to 0 dB, there will be no end threshold.
Under the parameter conguration menu use p and q to highlight “End Thre.” Press Enter to enter a value (Figure 3-21).
Press to exit.
Use t and u to adjust the position of the highlighted area. Use p and q to change the digits. After setting, press Enter to
conrm.
Note: The default setting is “03.00 dB.”
Delete File
This function deletes saved traces.
Under the parameter conguration menu use p and q to highlight “Delete File.” Press Enter to enter (Figure 3-22).
Press to exit.
Use p and q to choose the les to delete, and press Enter to conrm. Or use the arrow keys to choose “SelAll,” and press Enter
to select all les.
One or several les can be deleted at a time. Use the arrow keys to select “Delete.” Press Enter, and choose “Yes” to delete or
“No” to not delete. Choosing “Cancel” will exit the Delete menu.
Figure 3-21. End Threshold Conguration
Figure 3-22. Delete File