6 Description of the plasma care®
6.1 plasma care® components
The plasma care® is composed of the device, its docking station with the power supply and
the plasma care® spacer.
The docking station has two functions:
Safe placement and storage area for the plasma care®.
Recharging of the plasma care® via contactless, inductive energy transfer as soon as
the plasma care® is placed in the docking station.
As soons as the docking station is plugged in, the control-LED glows blue and the plasma
care® can be recharged.IIt takes about 3 hours to charge the plasma care® completely.
During the charging process the battery symbol of the plasma care® flashes red, yellow
and green.
The removable plasma source unit contains the plasma source, which is composed out of a
high voltage electrode, a dielectric barrier and a grounded electrode in the shape of a mesh
at the front.
6.2 plasma care® user interface
The user interface can be found on the upper side of the plasma care®. The interface allows
the following commands by pressing the touch-button:
Switching the plasma care® on / off by touching the button for 2 / 5 seconds
Starting and stopping a treatment by exerting pressure briefly.
The user interface displays the following information via differently colored LED-elements:
The system status, i.e. on / off, plasma source active
The battery charge condition
Warning and error notices
6.3 plasma care® spacer
The spacer (sterile single-use product) is attached to the plasma care® onto the plasma source
unit and has the following functions:
Designed as a sterile, single-use product to prevent cross-contamination.
plasma care®
docking station