Use Instructions

Instruction for Use
plasma care and plasma care spacer
IFU plasma care© and plasma care© spacer
Version: 0.1
Date: 26.05.2020
Copyright: terraplasma medical GmbH
Page 22/38
4.
As soon as the blue plasma ring glows permanently, the plasma care® is ready
for treatment.
Place the plasma care® gently onto the wound to be treated with the open side
of the spacer. The frame of the spacer should lie flat on the wound or the skin
such that a closed volume is created.
By pressing the on / off button briefly the plasma treatment is started.
Note: By briefly pressing the on / off button again the plasma treatment can be
stopped early.
5.
The four blue plasma ring segments represent a treatment duration of one minute;
each segment flashes for 15 seconds before being switched off.
As soon as the last plasma ring segment has been switched off, the plasma
source is also automatically switched off. The spacer can now be removed from
the device and be discarded. Alternatively, a second treatment of one minute
can be performed on the same area or on an adjacent area.
Note: If the plasma treatment is to be extended, the used spacer must remain
attached to the medical device. After a few seconds, the blue plasma ring re-
appears. To continue the treatment, press the on / off button briefly to restart
plasma generation. Three sequential one minute treatments may be given to a
wound area at any one time, however, a time interval of 12 hours must elapse
between two separate treatments. In total, 6 one minute treatments of the same
patient are possible using one spacer. The treatment of e.g. large wounds is pos-
sible by placing the spacer on six different, neighbouring areas of a large wound
and treating each area for one minute.
6.
By pressing the on / off button for an extended period of time (at least 5 sec-
onds), the plasma care® is switched off and the symbol in the middle of the user
interface turns grey.
When the plasma ring flashes red, this indicates that an invalid spacer is attached
to the plasma care®. The spacer must be exchanged, otherwise treatment is not
possible.
Note: In order to treat a new patient it is mandatory to exchange the spacer.