If the plasma ring glows red permanently, a device malfunction has occurred.
Note: Perform the steps described in chapter 10.4 Troubleshooting. If the plasma
ring continues to glow red permanently, please contact the customer service.
Ensure that the plasma care® and in particular the plasma source is not con-
taminated, smutched or damaged.
The treatment of the wound of a patient with the plasma care® may only be
performed once within a 12-hour window. The same section of a wound may
only be treated for a maximum duration of 3 minutes.
The treatment can be interrupted at any time by exerting extended pressure
onto the on / off button. Removal of the plasma care® spacer also leads to im-
mediate interruption of the treatment.
If the treatment cannot be stopped by either of these two options, the energy
supply of the plasma source can be interrupted by removal of the plasma
source.
Upon treatment with the plasma care® ozone is generated at low concentra-
tions. Treatments in the area of the head and particularly of the mouth should
be avoided.
8.2 Attaching and removing the spacer
8.2.1 Attaching the spacer
1. Open the sterile packaging of the spacer and place it on a solid surface such as a ta-
ble with the spacer still inside its sterile packaging.
2. Hold the plasma care® directly over the spacer as depicted. The clip of the spacer
needs to point towards the front of the medical device.
3. Now press the plasma care® onto the spacer until you hear a clicking sound. This indi-
cates that the spacer is securely attached to the plasma care®.
4. Lift the plasma care® with the attached spacer out of the packaging.
Clip
Press until it
“clicks”