If neccessary, remove rough dirt and deposits thoroughly. For this purpose, use a lint-free
cloth, warm water and a common detergent and wipe down the surface without touching
the plasma source or its mesh. Do not use scouring agents or aggressive detergents.
Allow the device to dry completely.
2. Disinfection:
The disinfection of the plasma care® must be performed using wipe disinfection.
Use a lint-free cloth or paper cloth for disinfection that is moistened with disinfectant.
Wipe down the plasma care® with this cloth as follows:
o Take particular care of the edge between the top and the body of the device.
Allow the disinfection agent to take effect according to the manufacturer’s instructions,
but at least 5 minutes.
Generally, disinfection wipes are recommended. Validated disinfection wipes are:
Bacillol® AF Tissues (BODE Chemie GmbH)
Pursept® A Xpress wipes (Schülke & Mayr GmbH)
ID032
Avoid excessive cleaning fluids, for example by frequent use of spray disin-
fectant, because the electronic components can be damaged. Cleaning
by submersion in a solution is prohibited.
9.1.2 Cleaning of the plasma source unit
If wound exudate or any other contaminant has deposited on the plasma source, the plasma
source unit needs to be cleaned as follows:
Detach the plasma source unit from the device:
1. Hold the plasma care® at the handle with one hand.
2. Grab the plasma source unit with the other hand at the edges on the side of the mesh.
3. Twist the plasma source unit a quarter of a turn counter-clockwise.
4. The plasma source unit can now be lifted off the plasma care®.