Cleaning the plasma source unit:
1. The disinfection of the plasma care® must be performed using wipe disinfection.
2. Use a lintless cloth or paper cloth for disinfection that is moistened very well with disin-
fectant; Bacillol® AF Tissues (BODE Chemie GmbH) or Pursept® A Xpress wipes (Schülke
& Mayr GmbH).
3. Dab the plasma source unit on the surface. Take care to not leave paper or cloth par-
ticles behind on the mesh of the plasma source. Avoid strong rubbing of the mesh.
4. Allow the disinfectant to take effect according to the manufacturer’s instructions.
5. Allow the plasma source unit to dry over night.
Reattach the plasma source unit to the device:
1. Hold the plasma care® at the handle with one hand again.
2. Grab the cleaned plasma source unit with the other hand at the edges on the side of
the mesh.
3. Position the plasma source unit in such a manner on the socket, that the large attach-
ment bolt fits into the large pocket and that the small attachment bolt fits into the small
pocket.
4. Twist the plasma source unit a quarter of a turn clockwise until it clicks into place. The
plasma source unit must then be flush with the body of the plasma care®.
5. The medical device is ready-for-use.