17 EMC
The following table shows the compliance with the respective standards regarding electro-
magnetic emission and immunity.
Phenomenon
EMC-Basic
standard
Immunity testing levels
Professionel healthcare
facilities
Home healthcare facilities
Electrostatic discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV Luft
High frequency electromag-
netic fields
IEC 61000-4-3
3 V/m
80 MHz up to 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz up to 2,7 GHz
80 % AM at 1 kHz
Power frequency magnetic
fiel
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Surge voltages
Wire against wire
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Conducted Interference in-
duced by High-Frequency
Fields
IEC 61000-4-6
3 V
0,15 MHz up to 80 MHz
6 V
at ISM-Frequencybands
between 0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 V
0,15 MHz up to 80 MHz
6 V
at ISM- and amateur radio fre-
quency bands between 0,15
MHz and 80 MHz
80 % AM at 1 kHz
Voltage drops
IEC 61000-4-11
0 % UT; 1/2 Periods
at 0, 45, 90, 135, 180, 225, 270 and 315 degrees
0 % UT; 1 Periode
and
70 % UT; 25/30 Periods
Single-phase: at 0 degrees
Voltage interruptions
IEC 61000-4-11
0 % UT; 250/300 Periods
Upon treatment with the plasma care® UV radiation is generated at very low
intensity. Avoid treatments at the eyes.