Use Instructions

Instruction for Use
plasma care and plasma care spacer
IFU plasma care© and plasma care© spacer
Version: 0.1
Date: 26.05.2020
Copyright: terraplasma medical GmbH
Page 37/38
The following table shows the compliance with the respective standards regarding immunity
against high frequency wireless communication devices.
Testfre-
quency
MHz
Frequency-
band
MHz
Radio service
Modulation
Maximal
power
W
Distance
m
Immunity
testing levels
V/m
385
380 to 390
TETRA 400
Puls modulation
18 Hz
1,8
0,3
27
450
430 to 470
GMRS 460
FRS 460
FM
± 5 kHz Hub
1 kHz Sinus
2
0,3
28
710
704 to 787
LTE Band 13, 17
Puls modulation
217 Hz
0,2
0,3
9
745
780
810
800 to 960
GSM 800/900
TETRA 800
IDEN 820
CDMA 850
LTE Band 5
Puls modulation
18 Hz
2
0,3
28
870
930
1720
1700 to 1990
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1,2,4,25
UMTS
Puls modulation
217 Hz
2
0,3
28
1845
1970
2450
2400 to 2570
Bluetooth
WLAN 802.11 b/g/n
RFID 2450
LTE Band 7
Puls modulation
217 Hz
2
0,3
28
5240
5100 to 5800
WLAN 802.11 a/n
Puls modulation
217 Hz
0,2
0,3
9
5500
5785
The following table shows the transmission- and receiving frequencies of the plasma care®
Receiving frequency
13,56 MHz
Preffered frequency
13,56 MHz
Receiver bandwidth
14 kHz
Transmission frequency
13,56 MHz
Transmitter bandwidth
14 kHz
Modulation of frequency
Amplitude modulation (ASK)
Effective radiated power
200 mW
18 FCC Certification
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired
operation.