Use Instructions

Patient Configurable Settings in Telemetry Setup (IIC)
11-12 Monitoring with the MX40 at the Information Center
Patient-Configurable Settings in Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Suppress
SpO
2
INOPs
with NBP
Enable/disable the
SpO
2
algorithm to
suppress sending
technical alarms from
the MX40 during an
NBP measurement for
60 seconds.
Warning
If you measure SpO
2
on a limb that has an
inflated NBP cuff, a
non-pulsatile SpO
2
technical alarm can
occur. If the monitor is
configured to suppress
this alarm, there can be
a delay of up to 60
seconds in indicating
critical patient status,
such as sudden pulse
loss or hypoxia.
enable
disable
enable
Pleth Wave
Enable/disable the
transmission of the
Pleth wave (and its
subsequent display) to
the Information Center.
For Continuous SpO
2
mode only.
enable
disable
Note When enabled, the Pleth
wave replaces the Vb wave in the
Patient Window during 6-lead
monitoring.
disable (Pleth is
not displayed.)
Pulse
Enable/disable display
of the Pulse rate at the
Information Center. For
Continuous SpO
2
mode
only.
enable
disable
disable (Pulse rate
is not displayed.)
SpO
2
Alarm
Turn SpO
2
alarms
on/off at the Information
Center and the MX40.
enable (on)
disable (off)
enable
Unit Settings
Change current settings
back to last saved
clinical unit settings.
(none)