Use Instructions
Unit Configurable Settings (IIC)
11-14 Monitoring with the MX40 at the Information Center
Unit Settings - Telemetry Setup
Control
Function
Settings
Factory
Default
Patient Type
Set patient type used for
SpO
2
alarm limits.
Adult
Pediatric
Adult
Telemetry/
Multi-Function
Button
Determine the Information
Center response when
Telemetry Button is
pressed.
Nurse Call - generate nurse
call alarm that can be
retrieved from Alarm Review
for later use.
Record - generate a recording
strip
Both - generate nurse call
alarm and recording strip
None
Nurse Call
Standby
Duration
Sets the standby duration
on the MX40 when
Standby is initiated from
the Information Center.
Note — When Standby
is initiated from the MX40,
the duration is always
Infinite.
Infinite
10 minutes
20 minutes
30 minutes
1 hour
2 hours
3 hours
4 hours
Infinite
Enable Remote
Suspend
Enable/disable alarm
pause/suspend at the
MX40.
enable
disable
disable
Suspend
Duration
Sets the alarm suspend
duration time for each
assigned device on the
Information Center.
1, 2, or 3 minutes
2 minutes
Battery Gauge
on Information
Center
Display/disable a battery
gauge for each assigned
device on the Information
Center.
enable
disable
enable (battery
gauge is
displayed)
Note — the
battery gauge is
always displayed
on the MX40.