Use Instructions

Unit Configurable Settings (IIC)
Monitoring with the MX40 at the Information Center 11-19
Unit Settings - NBP Setup
Control
Function
Settings
Factory
Default
Patient
Type
Set patient type used for NBP
alarm limits.
Adult
Pediatric
Adult
NBP Alarm
Set NBP alarm notification.
Systolic or Diastolic
Systolic
Diastolic
Mean
Off
Systolic or
Diastolic
Systolic
High
Increment/decrement NBP high
alarm limit by 5.
Limit Maximum is 260
Limit Minimum is 10
160 Adult
120 Pediatric
Systolic
Low
Increment/decrement NBP low
alarm limit by 5.
Limit Maximum is 260
Limit Minimum is 10
90 Adult
70 Pediatric
Diastolic
High
Increment/decrement NBP high
alarm limit by5.
Limit Maximum is 260
Limit Minimum is 10
90 Adult
70 Pediatric
Diastolic
Low
Increment/decrement NBP low
alarm limit by 5.
Limit Maximum is 260
Limit Minimum is 10
50 Adult
40 Pediatric
Mean High
Increment/decrement NBP high
alarm limit by 5.
Limit Maximum is 260
Limit Minimum is 10
110 Adult
90 Pediatric
Mean Low
Increment/decrement NBP low
alarm limit by 5.
Limit Maximum is 260
Limit Minimum is 10
60 Adult
50 Pediatric