Use Instructions

Global Settings (IIC iX)
Monitoring with the MX40 at the Information Center 11-23
Control
Function
Setting Choices
Factory
Default
Lead Placement
Determines the default lead
placement.
Standard
EASI
Standard
Screen-On Time
Determines how long the MX40
display is active by default.
1 min
2 min
5 min
15 min
30 min
1 min
Default Screen
Sets the default display screen.
2 Waves P (portrait)
1 Wave P (portrait)
2 Waves L (landscape)
Chest Diagram
2 Waves P
Wave 1-4
Selects the waveforms that will be
sent and stored. Wave 1 and
Wave 2 will always be Primary
and Secondary Lead respectively.
ECG
Pleth
Resp
Wave 1 and
Wave 2 will
always be
Primary and
Secondary
Lead
respectively.
ECG
ECG Management
Control
Function
Settings
Factory
Default
Primary Lead
Sets which lead will default to the
Primary Lead for ECG analysis.
1, II, III, aVR, aVF, aVL, V1-6,
V7-9, V3R - V5R
II
Secondary Lead
Sets which lead will default to the
Secondary Lead for ECG Multi
Analysis.
1, II, III, aVR, aVF, aVL, V1-6,
V7-9, V3R - V5R
V2
Va Lead
Sets the default Va lead label.
V1-6, V7-9, V3R-V5R
V2
Vb Lead
Sets the default Vb lead label.
V1-6, V7-9, V3R-V5R
V5
Filter
Sets the filter on the ECG wave
display.
0.5-40 Hz M
0.05-40 Hz ST
0.5-40 Hz M
Hexad (Va, Vb)
Sets the lead pairs for derived
12-lead ECG.
Off
V1, V3
V1, V4
V1, V5
V2, V4
V2, V5
V3, V5
V3, V6
Off