Physiologic Alarms
Alarms 5-13
Alarm Text
Priority
Condition
Source
* PVCs /MIN HIGH
Yellow
PVCs within one minute exceed by the
PVCs/min limit (xxx).
ST/AR Basic
& Enhanced
Arrhythmia
* R-ON-T PVCs
Yellow
For Heart Rate less than 100, a PVC with
R-R interval less than 1/3 the average
interval followed by a compensatory pause
of 1.25 x average R-R interval, or 2 such
Vs without a compensatory pause
occurring within 5 minutes of each other.
(When Heart Rate is greater than 100, 1/3
R-R interval is too short for detection.)
ST/AR
Enhanced
Arrhythmia
* RUN PVCs
Yellow
Run of PVCs greater than or equal to 2.
ST/AR
Enhanced
Arrhythmia
* SVT
Yellow
Run of SVPBs greater than or equal to
SVT Run limit and with SVT Heart Rate
greater than the SVT Heart Rate limit.
ST/AR
Enhanced
Arrhythmia
* VENT BIGEMINY
Yellow
A dominant rhythm of N, V, N, V (where
N= supraventricular beat, V=ventricular
beat).
ST/AR
Enhanced
Arrhythmia
* VENT RHYTHM
Yellow
A dominant rhythm of adjacent Vs greater
than Vent Rhythm limit and ventricular
Heart Rate less than V-Tach limit.
ST/AR
Enhanced
Arrhythmia
* VENT TRIGEMINY
Yellow
A dominant rhythm of N, N, V, N, N, V
(where N=supraventricular beat,
V=ventricular beat).
ST/AR
Enhanced
Arrhythmia