Application Notes (Temperature Differential)
When measuring temperature there may be applica-
tions when knowing the differential between two mea-
surements is required. This can be accompished by
following these steps.
Remove the test leads from the input jacks. Insert
the K-type thermocouple probe into the input jack
on the 270. Set the 270 to the desired TEMP
function.
Touch the temperature probe sensor to the device
under test (T1).
When the reading stabilizes, press the REL but-
ton. The display will read “0” and T1 will be
stored as the reference value that the next mea-
surement will be made relative to.
Touch the temperature probe sensor to the device
under test (T2).
The 270 will display the differential temperature
T2-T1.
Application Notes (AC Amps)
When measuring AC Amps of a motor there
are two types of measurements that can be
made, running current and in-rush or start-up
current. Start-up current will usually be much
higher than running current.
Set up the meter following the steps under
“Measurement Procedure” on page 17, and
then proceed with the following:
Clamp the meter around a single wire
and reconnect power to the device. Read
the current displayed on the meter. This
is the running current of the motor.
Disconnect power to the motor and put
the meter in PEAK HOLD mode.
Reconnect the power and read
the current displayed on the
meter. This is the in-rush or
start-up current
of the motor.