Instructions

Man_7005_1019 www.testec.de 7
4 Specifications
All specifications apply to the unit after a temperature stabilization time of 20
minutes over an ambient range of 25°C ± 5°C.
TT-SI 7005 – order number 15176
Bandwidth
60
MHz @ x
10
50 MHz @ x5
15 MHz @ x1
Rise Time
5,8
ns @ x
10
7 ns @ x5
23 ns @ x1
Attenuation
x
1
, x5, x
10
DC
-
Accuracy
±2%
CMRR
80
dB @ 60
Hz
60 dB @ 100 Hz
50
dB @ 1
MHz
Maximum Input Voltage @ x
1
(DC + AC peak)
±
3,5
V
Maximum Input Voltage @ x
5
(DC + AC peak)
±
18
V
Maximum Input Voltage @ x
10
(DC + AC peak)
±
35
V
Absolute Maximum Input Voltage
(each side to ground)
600
Vrms
Input Impedance
(differential)
1
MOhm //
2,5
pF
Input Impedance
(each side to
ground)
0,5
MOhm //
6
pF
Output Voltage Swing
±
4
V
using 1 M
Ω oscilloscope input
Offset (typical)
±5
mV
Noise (typical)
2
mVrms
Source Impedance
50
Ohm
Power Supply
9 V power adapter (included)
Weight
250
g
Dimensions
195 x 55 x 30
mm
BNC
cable length
100
cm
Input lead length
55
cm each
Operation Tempearure/Humidity
0°C to 50°C / 10% to 85%
RH
Storage Tempearure/Humidity
-
30
°C to
7
0°C / 10% to
90
%
RH
Safety Standard
IEC
6
1010
-
031:2015
CAT II
I
Pollution Degree
2
Specifications are subject to change without notice. To ensure the most current
version of this manual, please download the current version from our website:
www.testec.de