TESTEC High Voltage Probe TT-HVP 15 B
This high voltage probe is designed to prevent accidental shock to the operator when properly used.
This operating note must be read and understood prior to using the probe. Improper procedures or
incorrect analysis of the measurement situation can result in serious shock.
General Information
The Testec High Voltage Divider Probe is an accessory to be used with analog or digital oscilloscopes
having input resistance of 1 MΩ (± 1 %) and nominal input capacitance between 15 and 30 pF.
Specifications
Maximum Input Voltage Temperature Coefficient
DC Voltage: 10 kV Less than 200ppm/°C
AC Voltage: 7 kV RMS
PEAK Voltage: 20 kV
Compensation Range Accuracy
For input capacitance DC Volts: ±6%
of 15 pF to 30 pF
Rise Time
≤ 8,8 ns
Bandwidth Measurement Category
40 MHz (±3 dB) CAT 0
Other Characteristics
Input Impedance Operating Temperature
100 MΩ, 3 pF ± 0,5 pF 0 °C to +50 °C
Cable Length Storage Temperature
2 meter -20 °C to +70 °C
Altitude Humidity
Up to 4,600 meters (15.000 ft) Up to 80 % relative humidity at +40 °C
Specifications are subject to change without notice