User Manual
1
TT-SI 9010
70MHz HIGH VOLTAGE DIFFERENTIAL PROBE
USER’S MANUAL
This probe is in compliance with IEC-1010.1, IEC-1010.2-031
CAT. II, Pollution Degree 2.
1. Safety Terms and Symbols
Terms appear in this manual:
_______________________________________________
WARNING. Warning statements identify conditions or practices
that could result in injury or loss life.
_______________________________________________
CAUTION. Caution statements identify conditions or practices that
could result in damage to this product or other property.
Symbols appear on the product:
Danger Protective Attention
High Voltage (Earth) Terminal Refer to Manual
2. General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this probe
or any products that connected to it.
Observe Maximum Working Voltage
To avoid any injury, do not use the probe above 2500Vrms CAT II between each input lead
and earth or above 5000Vrms CAT II between two inputs. This voltage rating applies to both
of 1/100 & 1/1000 or 1/200 & 1/2000 settings.