6 Using the product
30
2. Move Joystick left/right to select the desired option: Auto-
scaling ( ), min. value ( ), temperature range ( ) or max.
value ( )
- The selected option has an orange border ( ).
3. With the selection of min. value, temperature range or max.
value: Move Joystick up/down to change the value(s).
- Changes are applied immediately so that the effects on the
display of the infrared image can be checked directly.
4. Close dialogue with [OK] or [Esc].
- The changes are saved.
6.1.4. Display...
Select display options
The presentation can be adapted by means of showing/hiding the
following elements: Crosshairs, temperature scale and quick select
buttons.
Hidden function keys can be shown again by pressing a quick
select button: The first press of the button shows the function keys
again. To trigger a function, the button must be pressed again.
1. [OK] | Display… | [OK].
- The Display options dialogue is opened.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border ( ).
3. Activate ( ) or deactivate ( ) option with [OK].
4. Confirm settings with the [Apply] quick select button.
or
Discard settings with [Esc].
6.1.5. Emissivity...
Emissivity
The emissivity describes the capability of a body to emit
electromagnetic radiation. This is material-specific and must be
adapted for correct measurement results.
Non-metals (paper, ceramic, gypsum, wood, paints and coatings),
plastics and food have high emissivity, which means that the
surface temperature can be easily measured using infrared.
Because of their low or non-uniform emissivity, bright metals and
metal oxides are only somewhat suited for infrared measurement.
You must anticipate highly inaccurate measurements. A remedy for
this is coatings that increase the emissivity, e.g. paint or emission