User Manual

5. Operation32
5
To sset AAuto HHold:
The Auto Hold function is only active on plug-in
measuring probes.
Configuration mode is opened, Auto Hold is lit.
1 Select the desired option (5, 10, 15, 20s) with and
confirm with :
· OFF: Readings are not recorded automatically.
· On: Once a stable reading is obtained (change in
reading <0.2 °C/0.4 °F in the set evaluation time), it
is recorded automatically.
OFF was selected:
º Continue with objective TO SET THE MAX./MIN. PRINT
FUNCTION
.
On was selected:
2 Use / to set the value for the evaluation time
(in s) and confirm with .
6
To sset tthe mmax./min.print ffunction:
Configuration mode is opened, MaxMin and are lit.
i Select the desired option with and confirm with
.
· On: Maximum and minimum values are printed out
as well when current or recorded readings are
printed.
· OFF: Maximum and minimum values are not printed
out as well when current or recorded readings are
printed.
7
To sset tthe ddate/time:
Configuration mode is opened, Year is lit.
1 Use / to set the current year and confirm
with .
2 Use / to set the other values for the month
(Month), day (Day) and time (Time) and confirm each
one with .