User Manual
9. Technical data38
9. Technical data
Characteristic Value
Parameters Temperature (°C/°F)
Measuring range NTC probe:
-50.0...+150.0°C / -58.0...+302.0°F
NTC high-temperature probe:
0.0...+275°C / 32.0...+527°F
Resolution 0.1°C / 0.1°F
Accuracy NTC probe:
(±1 Digit) ±0.2°C / ±0.4°F (-20.0...+80.0°C / -4.0...+176.0°F)
±0.3°C / ±0.6°F (rest of range)
NTC high-temperature probe:
±0.2°C / ±0.4°F (0.0...+80.0°C / 32.0...+176.0°F)
±0.3°C / ±0.6°F (rest of range)
Sensor connections 1x Mini-DIN socket for NTC temperature probe,
radio module (accessory part)
Measuring rate 2/s
Operating temperature range -20 ... +50°C / -4 ... +122°F
Storage temperature -40 ... +70°C / -40 ... +158°F
Voltage supply 1x 9 V monobloc battery/rech. battery
Running time (display with probe connected: approx. 200h / approx. 68h,
lighting off / on) with radio probe: approx. 45h / 33h
Protection class with TopSafe (accessory part) and probe connected: IP65
EC Directive 89/336/EEC
Warranty 2 years
With TopSafe and the following probes, this product complies with guidelines in
accordance with the EN 13485 standard:
Part no. Measuring range
0613 1001 -50...+275°C / -58.0...+527°F
0613 1212 -50...+150°C / -58.0...+302°F
0613 1712 -50...+150°C / -58.0...+302°F
0613 2211 -50...+150°C / -58.0...+302°F
0613 2411 -50...+150°C / -58.0...+302°F
0613 3211 -50...+140°C / -58.0...+284°F
Suitability: S, T (storage, transport)
Environment: E (transportable thermometer)
Accuracy class: 0.5
Measurement range: see table above
According to EN 13485, the measuring instruments should be checked and
calibrated regularly under the terms of EN 13486 (Recommended: Yearly).
Contact us for more information: www.testo.com