YES was selected:
- Automatic probe detection is started. Auto flashes
while the instrument looks for a radio probe that is
switched on.
- Once a radio probe is found, the probe ID is
displayed. If no probe is found, NONE lights up.
Possible reasons why probes are not found:
· The radio probe is not switched on or the
battery of the radio probe is spent.
· The radio probe is outside the range of the
measuring instrument.
· Sources of interference are influencing the radio
transmission (e.g. reinforced concrete, metal
objects, walls or other barriers between
transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
i If necessary, rectify the possible causes for the
disruption to the radio transmission and start
automatic probe detection again with .
- If further wireless probes are within reception range,
the probe ID of a different wireless probe may be
displayed.
i If necessary: switch off other wireless probes or
remove from the reception range, and start
automatic probe detection again with .
2 Press to change to the next function.
4
To sset AAuto OOff:
Configuration mode is opened, Auto Off is lit.
i Select the desired option with and confirm with
:
· On: The measuring instrument switches off
automatically if no button is pressed for 10 min.
Exception: A recorded reading is shown on the
display (Hold or Auto Hold is lit).
· OFF: The measuring instrument does not switch
itself off automatically.