User Manual
38
9. Technical data
Characteristic Value
Parameters Temperature (°C/°F / °R)
Measuring range -50.0...+400°C / -58.0...+752.0°F / -40.0...+320°R
Resolution 0.1°C / 0.1°F / 0.1°R (-50.0...+199.9°C / -
58.0...+391.8°F / -40.0...+159.9 °R)
1°C / 1°F / 1°R (rest of range)
Accuracy ±0.3°C / ±0.6°F / ±0.2°R (-20.0...+70.0°C /
(±1 Digit) -4.0...+158.0°F / -16.0 ... +56.0°R)
±0.7°C+0,5% o.r. / ±1.3°F+0,5% o.r. / ±0.6°R + 0,5%
o.r. (rest of range)
Probe connections 1x Omega TC socket for temperature probe
type T (Cu-CuNi), radio module (accessory part)
Measuring rate 2/s
Operating temperature range -20 ... +50°C / -4 ... +122°F / -16 ... +40°R
Storage temperature -40 ... +70°C / -40 ... +158°F / -32 ... +56°R
Voltage supply 1x 9 V monobloc battery/rech. battery
Running time (display with probe connected: approx. 200h / approx. 68h,
lighting off / on) with radio probe: approx. 45h / 33h
Protection class with TopSafe (accessory part) and probe connected: IP65
EC Directive 2004/108/EC
Warranty 2 years
With TopSafe and the following probes, this product complies with guidelines in
accordance with the EN 13485 standard:
Part no. Measuring range
0613 1001 -50...+275°C / -58.0...+527°F
0603 1293 -50...+350°C / -58.0...+662°F
0603 1793 -50...+350°C / -58.0...+662°F
0603 2192 -50...+350°C / -58.0...+662°F
0603 2492 -50...+350°C / -58.0...+662°F
0603 3292 -50...+350°C / -58.0...+662°F
Suitability: S, T (storage, transport)
Environment: E (transportable thermometer)
Accuracy class: 0.5
Measurement range: see table above
According to EN 13485, the measuring instruments should be checked and
calibrated regularly under the terms of EN 13486 (Recommended: Yearly).
Contact us for more information: www.testo.com
9. Technical data