Product Overview

testo 735
Probes
Probe type
Flat head surface probe with
telescopic handle max. 680 mm for
measurements at hard-to-access
points, TC Type K, Fixed cable 1.6
m (correspondingly shorter when
telescope extended)
Magnetic probe, adhesive force
approx. 20 N, with magnets, for
measurements on metal surfaces,
TC Type K, Fixed cable
Magnetic probe, adhesive force
approx. 10 N, with magnets, for
higher temp., for measurements on
metal surfaces, TC Type K, Fixed
cable
Waterproof surface probe with
widened measurement tip for flat
surfaces, T/C Type K, Fixed cable
Pipe wrap probe with velcro strip;
for temperature measurement on
pipes with diameter up to max. 120
mm; Tmax. +120 °C; TC Type K ,
Fixed cable
Pipe wrap probe for pipe diameter
5 to 65 mm, with exchangeable
measuring head. Meas. range
short-term up to +280°C, TC Type
K, Fixed cable
Spare meas. head for pipe wrap
probe, TC Type K
Clamp probe for measurements on
pipes, pipe diameter 15 to 25 mm
(max. 1"), meas. range short-term
up to +130°C, TC Type K, Fixed
cable
Waterproof surface probe with
widened measurement tip for flat
surfaces, T/C Type T, Fixed cable
1.2 m
Robust, waterproof Pt100
immersion/penetration probe, Fixed
cable
Highly accurate Pt100
immersion/penetration probe incl.
calibration protocol (test points 0
°C and +157 °C), Fixed cable
Dimensions
Probe shaft/probe shaft tip
Measuring
range
-50 to +250 °C
-50 to +170 °C
-50 to +400 °C
-60 to +400 °C
-50 to +120 °C
-60 to +130 °C
-60 to +130 °C
-50 to +100 °C
-50 to +350 °C
-50 to +400 °C
-40 to +300 °C
Accuracy
Class 2
2)
Class 2
2)
Class 2
2)
Class 2
2)
Class 1
2)
Class 2
2)
Class 2
2)
Class 2
2)
±0.2 °C (-20 to +70 °C)
Class 1 (remaining range)
2)
Class A (-50 to +300 °C), Class
B (remaining range)
1)
±0.05 °C (0 to +100 °C)
±(0.05 °C +0.05% of mv)
(remaining range)
t
99
3 s
150 s
30 s
90 s
5 s
5 s
5 s
30 s
12 s
60 s
Part no.
0602 2394
0602 4792
0602 4892
0602 1993
0628 0020
0602 4592
0602 0092
0602 4692
0603 1993
0609 1273
0614 0235
Surface probes
Immers./penetr. probes
680 mm
35 mm
75 mm
115 mm
395 mm
35 mm
112 mm
114 mm
295 mm
12 mm
20 mm
15 mm
50 mm
50 mm
Ø 5 mm
Ø 5 mm
Ø 5 mm
Ø 4 mm
Ø 25 mm
Ø 20 mm
Ø 21 mm
Ø 6 mm
Ø 6 mm
Ø 3.7 mm
1) According to standard 60751, the accuracies of Class A and B refer to -200 to +600 °C (Pt100)
2) Accor
ding to standard EN 60584-2, the accuracy of Class 1 r
efers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
We measure it.
Information on surface measurement:
The response times t
99
stated are measured on ground steel or aluminium plates at +60 °C.
The stated accuracies are sensor accuracies.
The accuracy in your application is dependent on the surface structure (roughness), material of the measurement object (heat capacity and heat transfer), as wel as sensor accuracy.
Testo creates a corresponding calibration certificate for the deviations of your measurement system in your application. For this purpose, Testo uses a surface test bench developed in
cooperation with the PTB (Physikalisch Technische Bundesanstalt).
testo-735-P06 03.12.2012 08:02 Seite 6