6.2.4 Cyclical Print (735-1 only)
The Cyclical Print function can be activated/deactivated. A measuring program
for cyclical printing can be programmed. This enables readings (up to 999) to
be printed in a defined measuring cycle (min. 1min). The readings are sent to a
Testo printer.
²
Activating ccyclical pprinting/programming aa mmeasuring pprogram:
The main menu is open, Menu is displayed.
1 cyc.Print ¬ .
2 Press / to select Off (deactivated) or On (activated) and confirm
with .
The measuring cycle is set in the order: minutes/hours.
3 Press / to set the measuring cycle in minutes and confirm with
.
4 Perform the setting for hours as described in step 3.
5 Press / to set the number of readings and confirm with .
- The instrument returns to measurement view.
6.2.5 Alarm
The alarm thresholds can be set. The default settings for the alarm thresholds
are the limit values for the measuring range.
If an alarm threshold is exceeded or undershot during a measurement, an alarm
tone is emitted.
testo 735-2 only: The alarm thresholds are related to the location. They are
only applied to the location that was activated in the setting.
²
Setting aalarm tthresholds:
The main menu is open, Menu is displayed.
testo 735-2 only:
i Activate the location for which the settings are to apply.
1 Alarm ¬ .
2 Press / to select Max (upper alarm threshold) or Min (lower alarm
threshold) and confirm with .
3 Press / to set the value and confirm with .
OK
OK
OK
OK
OK
OK
OK
6. Setting the instrument48