User Manual

8 Alarms
18
Also send report via e-mail: as well as saving reports under Generated
reports, these can also be sent as e-mails. Click on the checkbox to open
the input screen for e-mail addresses.
3. Click on the Create an automatic report button.
The first report will be created on the following day.
Generated reports
A summary of the reports already generated is displayed.
Click on the icon to display more information.
Click on the Download button to download a report.
Click on the Edit this report series button to display and edit the settings.
Report settings
Automatic reports which have already been created are displayed in a table.
Click on the Actions button and then on Edit to display and edit the settings.
Click on the Actions button and then on Delete to delete the automatic
report.
8 Alarms
8.1 Alarm list
Display of alarms
A summary of all triggered alarms and system warnings is displayed. Unread
alarms and system warnings are shown in bold.
The display can be filtered according to the following characteristics:
By measuring point group/measuring point: click on the checkbox in front of
the measuring point group/measuring point.
By date/time: click on the start date/end date and select the start date/start
time and end date/end time.
Detailed information on alarms
Click on the icon to display more information.
On displaying the detailed information, the alarm message/system warning is
marked as "read" and the alarm counter is marked down.
Click on the Mark all as read button to mark all alerts as "read".
8.2 Alarm settings
Creating and displaying alarm settings
Click on the + New alarm setting button to create a new alarm setting.
Existing alarm settings are displayed below the button.
Click on the title of an alarm setting to display it.
Configuring and editing a displayed alarm setting
The following settings can be defined and edited:
Title: designation of the alarm setting (required field).
Validity range: measuring point group/measuring point which is to be
monitored. Click on the checkbox in front of the measuring point
group/measuring point.
Lower limit value, Upper limit value: values which are to be monitored.
Channel alarms: alarms in the event of a defective sensor.
E-mail recipient: addressees who are informed when an alarm occurs. Click
on the checkbox in front of the recipient or enter the name and e-mail
address of other recipients and click on the + Add button.