User's Manual

4 Product use
9
4.2 Reading out data
1
Remove the data logger and
immediately press the Start/Check
button. While the button is pressed,
the data logger indicates whether the
limit values were adhered to/violated.
- LED lights up green:
limit values adhered to
- LED lights up red:
limit values violated
CAUTION
If too much time elapses between
removing the data logger and
pressing the button, this can lead
to a limit value violation due to a
deviating ambient temperature.
2
Pressing the button establishes a
connection to the readout unit (order
no. 0572 2802). The measurement
data is transmitted to the database
via the readout unit.
CAUTION
If no connection can be
established, the data logger
spends 60 seconds attempting to
set up the connection. If no data
transfer can be carried out within
this time, the data logger resumes
its measurement. If this happens,
please repeat the readout
process.
3 Following successful data transfer,
the LEDs flash red and green at one-
second intervals, continuously.